OTHER NON-LAPAROSCOPIC
PROCEDURES:
1. Installation should already be completed according to the
installation instructions.
2. Be familiar with all operating controls as described in
Description of Switches, Control Buttons, and Indicators.
3. Refer to Figure 22 or 23 for set-up.
4. Connect SAFEGUARD BLUE
Filter with Built-in Fluid Trap to CRYSTAL VISION
5. Connect Laser Hand piece Accessory or ESU Handpiece
Accessory to the SAFEGUARD BLUE
ULPA Filter with Built-in Fluid Trap .
6. Connect ESU sensor to ESU port of the CRYSTAL
®
VISION
or Laser Sensor Cable to Sensor Assembly and
the CRYSTAL VISION®.
7. Plug CRYSTAL VISION
8. Turn ON POWER switch on back panel.
9. Push OPEN (Orange) button. Indicator should illuminate.
10. Adjust FLOW SETTING for the desired flow.
11. Push MANUAL button and observe that desired flow is
shown on the digital FLOW meter display
12. Adjust TIME SETTING for the desired amount of time
that the CRYSTAL VISION® runs after the smoke
producing equipment is deactivated.
IF SMOKE PERSISTS DURING THE
SURGICAL PROCEDURE, TRY THE
FOLLOWING:
13. Increase the FLOW SETTING and/or the TIME
SETTING if smoke is not eliminated from the surgical
site.
14. CHANGE FILTER light may indicate a partially
obstructed speculum port or tubing. Check and clean
them, if necessary. The SAFEGUARD BLUE®
Hydrophobic ULPA Filter with Built-in Fluid Trap may
also need to be changed.
15. OCCLUSION alarm indicates an obstructed smoke
collection nozzle, kinked or obstructed smoke tubing.
Check and clean as necessary. It may also mean that
the SAFEGUARD BLUE® Hydrophobic ULPA
Filter with Built-in Fluid Trap has become extremely
filled with smoke particles. Replace it, if required.
IF SMOKE PERSISTS WHEN THE
PUMP STOPS BE SURE THAT:
16. Tubing is not crimped
®
17. SAFEGUARD BLUE
Built-in Fluid Trap is clean, may need to be replaced if it
is clogged
18. Be sure there are no leaks in the tubing, or instruments:
19. Increase the TIME SETTING if there are no leaks.
®
Hydrophobic ULPA
®
Hydrophobic
®
into power outlet.
Hydrophobic ULPA Filter with
®
.
Figure 22: Set up for using CRYSTAL VISION® with an ESU
system during an open procedure.
Figure 23: Set-up for using a CRYSTAL VISION
hand piece.
19
®