7.13 Counting
Purpose:
Use this application to determine the number of parts of approximately equal weight .
The weight of a counted reference sample is calculated and then the objects with an
unknown piece count are weighed . The balance displays the number of parts and the
piece weight .
Minimize counting errors:
Ensure that the weight of individual parts is evenly distributed .
Larger reference piece counts lead to a higher degree of accuracy .
Switch on Auto Reference Updating .
Select the Menu key in any application .
t
Select the
t
g
Counting
y The
The piece count of the reference sample is displayed under
d=0.01 g
V
w
5
pcs
To set the reference sample quantity, select the gray button .
t
Reference
5
pcs
l
Counting
y The
Optimizing
Enter the reference piece count on the left .
t
Select a value or select
Automatic
Select the accuracy of the weight value during counting in the center .
t
y Activate "Automatic Reference Sample Updating" on the right .
This "updating" optimizes the counting accuracy . The average piece weight is
Off
recalculated every time the reference sample is updated automatically . Since the
new parts enlarge the basis for the calculation, this increases the counting
accuracy of the reference and with it the result too .
Select
l
t
Select
A
t
Place a container on the weighing pan .
t
®
, Practum
®
Counting
symbol in the menu .
application appears .
settings window appears .
t
, enter the desired value, and confirm with
to confirm .
to zero the balance if necessary .
Reference
.
l
.

Advertisement