3. Connect the equipment as shown in Figure 3-3. The leads should be
4. Set the Agilent 432A range switch to coarse zero and adjust the front
5. Fine zero the Agilent 432A on the most sensitive range, then set the
Ensure that the DVM input leads are isolated from chassis ground
Note
when performing the next step.
6. Set up the DVM to measure microvolts.
7. Connect the positive and negative input leads, respectively, to the
8. Observe the reading on the DVM. If less than 400 µV, proceed with the
11.Record the reading on the DVM to two decimal places. This is V
12.Reconnect the negative lead to the V
13.Calculate the required value of V
Agilent E4418B/E4419B Service Guide
isolated from ground. Ensure that the power reference oscillator is off.
Ensure that both the power meter under test and the Agilent 432A
have been powered on for at least 30 minutes before proceeding to the
next step.
panel coarse zero control to obtain a zero meter indication.
range switch to 1 mW.
V
and V
connectors on the rear panel of the Agilent 432A.
comp
RF
next step. If 400 µV or greater, press and hold the Agilent 432A fine
zero switch and adjust the coarse zero control so that the DVM
indicates 200 µV or less. Then release the fine zero switch and proceed
to the next step.
9. Round off the DVM reading to the nearest microvolt and record
this value as V
.
0
• V
________________µV
0
• Disconnect the DVM negative input lead from the V
the test power meter and reconnect it to chassis ground.
10. Turn the power reference on by pressing
Zero
,
Power Ref On
Cal
• V
_____________V
comp
the Agilent 432A. The DVM is now set up to measure V
represents the power reference oscillator output level.
V
=
V
+
V
V
1
comp
0
Power Reference Oscillator Level Adjustment
.
connector on the rear panel of
RF
using equation 2.
1
3
2
(
)
(
) CalibrationFactor
(
10
4R
comp
Adjustments
connector on
RF
.
comp
which
1
)