Agilent Technologies 4418B Service Manual page 46

Power meter
Table of Contents

Advertisement

Performance Tests
Instrument Accuracy Test
Table 2-3: Instrument Accuracy Results (For Indication Only)
Range
Calibrator
Setting
3 µW
10 µW
30 µW
100 µW
300 µW
1 mW
3 mW
10 mW
30 mW
100 mW
1. These performance limits are determined by the zero set specification of the power sensor
used in the measurement plus the measurement noise.
Note
Note
2-10
Channel A
1
Min
Actual Results
3.100 µW
_______________
9.900 µW
_______________
31.40 µW
_______________
99.50 µW
_______________
314.00 µW
_______________
0.995 mW
_______________
3.141 mW
_______________
9.984 mW
_______________
31.63 mW
_______________
100.9 mW
_______________
The nominal outputs for the 3 µW, 30 µW, 300 µW, 3 mW, 10 mW,
30 mW and, 100 mW settings are 3.16 µW, 31.6 µW, 316 µW,
3.156 mW, 10.03 mW, 31.78 mW, and 101.3 mW respectively.
It is not necessary to check instrument accuracy in dBm. The
power meter uses the same internal circuitry to measure power
and mathematically converts watts to dBm.
Channel B
Actual Results
(Agilent E4419
B only)
3.230 µW
_______________
10.10 µW
_______________
31.80 µW
_______________
100.5 µW
_______________
318.00 µW
_______________
_______________
1.005 mW
_______________
3.171 mW
_______________
10.08 mW
_______________
31.94 mW
_______________
101.8 mW
Agilent E4418B/E4419B Service Guide