Performance Tests
Instrument Accuracy Test
14. Set the DC calibrator output to the values shown in Table 2-4. For
each setting, set the filter length of the power meter to the
corresponding value, and verify that the power meter reading is
within the limits shown. Refer to step 10 for details on how to
setup the filter length
.
Table 2-4: Instrument Accuracy Results
Power
Meter Meas
Filter Unc
Setting
256
0.028 uW
256
0.027 uW
256
0.027 uW
64
0.073 uW
64
0.104 uW
16
0.2 uW
16
0.9 uW
16
1.6 uW
16
4.8 uW
16
18.5 uW
2-14
Lower
Ch A
Ch B
Limit
Resul
t
Result
3.0977uW
9.904 uW
31.428uW
99.49uW
314.72uW
0.9954mW
3.1477mW
9.954mW
31.477mW
99.54mW
Agilent E4418B/E4419B Service Guide
GuardBand GuardBand
Upper
Lower
Limit
Limit
(in µW)
3.2269uW
3.1257
10.096uW
9.931
31.819uW
31.455
100.51uW
99.563
317.74uW
314.824
1.0047mW
995.6
3.1769mW
3148.6
10.046mW
9955.6
31.769mW
31481.8
100.46mW
99558.5
Upper
Limit
(in µW)
3.1989
10.069
31.792
100.437
317.636
1004.5
3176
10044.4
31764.2
100441.5