Test Procedure 2
Note
Equipment
Test Setup
Agilent E4418B/E4419B Service Guide
This test procedure makes use of a 11683A Range Calibrator fitted
with option H01. The measurement uncertainty of this test
procedure is significantly lower than the previously described
method (refer to Appendix B for a description of the measurement
uncertainty analysis). As such, this test procedure should be used
to ensure the power meter meets it's published specification.
•
Range Calibrator: Agilent 11683A Option H01
•
Power Sensor Cable: Agilent 11730A
•
DC Range Calibrator: Datron 4000A
Power Meter
CHANNEL A
Figure 2-3: Instrument Accuracy (Average Path) Test Setup
Range
Power sensor cable
Calibrator
Performance Tests
Instrument Accuracy Test
DC Calibrator
2-11