Agilent Technologies 4418B Service Manual page 45

Power meter
Table of Contents

Advertisement

Note
Agilent E4418B/E4419B Service Guide
When switching the range calibrator to STANDBY, allow enough
time for the range calibrator to settle to its zero value before
attempting to zero the power meter. This settling would appear on
the power meter display as downward drift. When the drift has
reached minimum, (typically less than 60 seconds), the range
calibrator is settled.
6. On the power meter press
Filter Filter On Mode Man Length
,
length to 512 in the pop up window then press
7. On the power meter press
seconds for the wait symbol to disappear. Verify that the display
reads 0 ±0.05 µW.
8. Set the range calibrator's FUNCTION switch to CALIBRATE.
9. Set the range calibrator's RANGE switch to 1 mW.
10. Press
to calibrate the power meter.
Cal
11. Set the range calibrator's RANGE switch to the positions shown in
Table 2-3 on page 2-10. For each setting, verify that the power
meter's reading is within the limits shown.
Instrument Accuracy Test
System
,
Input Settings
Inputs
,
,
, and set the filter
Zero
,
. Wait approximately 10
Zero
Cal
Performance Tests
,
,
More
.
Enter
2-9

Advertisement

Table of Contents
loading

Table of Contents