Agilent Technologies 4418B Service Manual page 124

Power meter
Table of Contents

Advertisement

Troubleshooting
Introduction
Select
Fail O/P High
DVM reading
Yes
0.0V+/-0.5V?
No Fault found with Upper
Window TTL Over Limit
detection circuit - check user
application/implementation
6-6
A
Select
Preset/local
Confirm
Lower Window - using
Meas/Setup
Limits
Limits On
TTL Output
TTL Output On, Fail O/P Low
DVM reading
Yes
5.0V+/-0.5V?
No
A2 Processor Assy
Defective
End
No
Power down meter.
Disconnect ac power and battery if fitted.
Remove A2 Processor assy, as directed on page
5-9, to access the Comms Assembly connector.
Use DVM to check resistance between J4 pin 42
and Rmt I/O connector pin 4.
DVM reading
Yes
450Ω +/- 10%
Agilent E4418B/E4419B Service Guide
No
A4 Comms Assy
Defective