3-9-3 List of Functions

For details of each function, see the following Table 3-9-3.
Table 3-9-3 Various Functions in Function Selecting Mode
Function Name
F-01
Original Value
F-02
Actual Load Calibration
F-03
Sensitivity Registering
Calibration
Initial value:2.000 [mV/V]
4000 Display
Key Operation and WGI-400A
Movement in Function Setting Mode
"......" is displayed.
Press the SET key and the input value at
this time is displayed in mV/V.
Note: Before activating the calibration
function, always conduct digital
Zero function with no load applied
to the transducer.
Apply load to the transducer
Input a desired numeric value (load) to
be indicated and press the SET key.
Set a decimal point and press the SET
key. Then, the calibration value is
updated.
Remove the load.
Set rated output of the transducer in
mV/V and press the SET key.
Input the desired indicated value (rated
capacity) and press the SET key.
Set a decimal point and press the SET
key. Then, the calibration value is
updated.
Setting available range:
Less than 9999 per 0.1 mV/V
20
Functional Description
• A function that displays the
raw output value of the
transducer.
• A function that applies the
actual load to the transducer
and sets the indicated value
of that time.
• For details, see "3-4-1
Actual Load Calibration."
• A function that conducts
calibration by registering
the rated output and rated
capacity of the transducer
without applying the actual
load.
• For
details,
see
Sensitivity
Registering
Calibration Function"