3-6 PATTERN FUNCTION

High and low limit values are stored in 4 sets of operation patterns (0 to 3). They can be changed from external
control input terminal or by key operation.
1) Writing to Operation Patterns of High/Low Limit Values
When pattern No. (0 to 3) is displayed on the Status Indicator ([STATUS]) on the front panel, the preset
high or low limit values are stored as patterns 0 to 3.
Reference: Once stored set value is maintained even after the power is cut off.
2) Selecting Operation Pattern
• Operation pattern can be selected by setting function selecting mode 'F-14.'
For operation procedures, see function selecting mode 'F-14.'
Note: Selecting operation pattern is enabled when 'Pattern Select Command 1' or 'Pattern Select
Command 2" of external control input is set to OFF.
• Operation pattern is selected by 'Pattern Select Command 1' or 'Pattern Select Command 2' of the external
control input.
Operation pattern 0
Operation pattern 1
Operation pattern 2
Operation Pattern 3
Reference: Operation pattern is selected by giving priority to the external control input. When it is
selected by the external control input, set value of Function 'F-14' is cancelled to have the
initial value '0.'

3-7 LEVEL TEST FUNCTION

Level test function is used for setting the indication of the WGI-400A to a predetermined indicated value. It is
used for adjusting the external connected devices or testing the output of the WGI-400A.
1) Setting Function
• Additional Function
0 disabled:
1 enabled:
• Setting level test value
Level test value is set within the range ±9999 and the initial value shall be 1000.
For how to set the level test value, see function selecting mode 'F-25.'
2) Operation
Set value is continuously displayed while the level test command of the control input is set to close (ON).
Pattern Select Command 1
Displays the indicated value irrespective of the measured value.
Displays the indicated value by adding the set value to measured value.
Open (OFF)
Close (ON)
Open (OFF)
Close (ON)
17
Pattern Select Command 2
Open (OFF)
Open (OFF)
Close (ON)
Close (ON)