KYOWA WGI-400A Instruction Manual page 25

Instrumentation amplifier with multipurpose indicator
Hide thumbs Also See for WGI-400A:
Table of Contents

Advertisement

Table 3-9-3 Various Functions in Function Selecting Mode
Function Name
F-04
Numeric Value Registering
Calibration
(Calibration Zero
Registration
F-05
Self-Check
F-06
Analog Output
Initial value
'0':
Zero
'4000': Full scale
.
Key Operation and WGI-400A
Movement in Function Setting Mode
Displays calibration Zero value.
Changes the numeric value by key
operation.
Input the initial value of the transducer.
Press the SET key to determine the
indicated value.
"......" is displayed.
Press the SET key and all the indicator
LEDs light up to conduct self-check.
The self-check results are indicated as
described at the right.
● If faulty self-check results appear on
the indicator, contact KYOWA or our
representatives.
• Select output
0: Voltage output
1: Current output
Select and press the SET key.
'◦' is displayed on the Status Indicator.
Set the indicated value to have the
output to '0 V or 4 mA' and press the
SET key.
'F' is displayed on the Status Indicator.
Set the indicated value to have the
output to '10 V or 20 mA' and press the
SET key.
Setting range: ±9999
21
Functional Description
A function that re-sets the
Zero value (a reference
value) when calibration.
For
details,
see
Numeric Value Registering
Calibration."
• A function that self-checks
the internal memory and
displays the checked results.
Good: Normal
E-01: RAM failure
E-02: Flash memory
failure
A function that sets and
defines Zero and full scale
with the indicated value.
Select the desired analog
output.
• If the analog output exceeds
the full scale value by
approximately 5%, about 12
V is output. If it lowers to -
full scale, -12 V is output.
Current output is outputted
in
the
same
approximately 12 mA and
0 mA.
(2/5)