Table 3-8-3 Various Functions in Function Selecting Mode
Function Name
F-07
Smoothing
Initial value:
'0.0'
Moving average: None
Minimum scale: 1
F-08
Additional Value
Initial value: '0'
Key Operation and WGI-400A
Movement in Function Setting Mode
Displays the following items 1) and 2).
1)
Number of moving
average times
O:
None
1:
2 times
2:
4 times
3:
8 times
4:
16 times
5:
32 times
6:
64 times
7:
128 times
Select and change values in items 1) and 2)
and press the SET key.
Note: When moving average times is set,
response of the analog output is
deteriorated.
• Set the desired additional value and
press the SET key.
● Setting range: ±9999
2)
Minimum scale
0:
1
1:
2
2:
5
3:
10
4:
20
5:
50
6:
100
22
Functional Description
A function that stables the
indication by smoothing the
varying signals from the
transducer.
Number of moving average
times
The number of moving
average times to average
and display the measured
value.
Minimum scale
Minimum unit that varies
the indicated value.
To set the number of
display times relevant for
the WGI-300, select '4' in
item 1) and set '16.'
• A function that adds the
desired numeric value to the
measured value for display.
• Conduct
digital
function with the preset
additional value. Then, the
obtained value is displayed
on the indicator.