Table 3-9-3 Various Functions in Function Selecting Mode
Function Name
F-09
Auto Zero Compensation
Initial value: '0'
'0.0'
'0'
F-10
Hysteresis
Initial value: '0'
'0'
F-12
Initialize
F-13
BV Select
Initial value: '0'
Key Operation and WGI-400A
Movement in Function Setting Mode
•
'μ' is displayed on the Status Indicator.
Set Zero compensation range and press
the SET key.
•
't' is displayed on the Status Indicator.
Set the judging time and press the SET
key.
●
Zero Compensation range
●
Judging time:
•
'2' is displayed on the Status Indicator
(ZERO).
Set Zero near Zero range and press the
SET key.
●
Zero near Zero range: 0 to 9
•
'1' is displayed on the Status Indicator.
Set hysteresis of the high limit
comparator and press the SET key.
•
'2' is displayed on the Status Indicator.
Set
hysteresis
of
comparator and press the SET key.
●
Setting range: 0 to 9999
• Initialize
0: Not initialized
1: Initialized
Press the SET key. Then, the set value
returns to factory default setting.
● Calibration value and BV set value are
not initialized.
• Set BV set value. 0: 2 V
Select either of the above volts and
press the SET key.
: ±9999
0.0 to 9.9 (sec)
the
low
limit
1: 4 V
23
Functional Description
•
Zero tracking
When
it
passes
judging
time
measured value is within
the Zero compensation
range,
this
activates
digital
function
automatically Zeros the
indicated value.
•
Zero near Zero range
When the measured value
is within the Zero near
Zero
range
setting
range),
function
automatically
Zeros the indicated value.
• A function that sets the
hysteresis of the high and
low
limit
comparators
(Comparators 1 and 2).
• A function that returns the
set values to factory default.
• A function that selects and
changes
the
excitation voltage.
(4/5)
the
and
function
Zero
and
(Within
this
bridge