7.9. Display errors

The program can read out and display the error messages when the
serial port is open. When the error occurs, the error message and the
error code are visible.
The figure shows how visible the "Output range error" status.
When the critical errors occur (Error code = 1), must call the
manufacturer.
When the next errors occur:
Input Range Error (Error code = 4): The input signal is too high
(input current / voltage). Please check the transmitter which connects
to the input and the input ranges.
Please read out the settings from the instruments, and check them. If
you find bad setting, please fix it, and download new settings to the
instrument.
Output Range Error (Error code = 5): The output signal is too high
(output current / voltage). Please check the transmitter which connects
to the input and the input/output ranges.
Please read out the settings from the instruments, and check them. If
you find bad setting, please fix it, and download new settings to the
instrument.