Adjustment and Calibration
Figure 4.1 Results Calibration Check Screen
4.3

Calibration File Installation

Purpose
The ATLAS software requires fixed optics head data to acquire exams. This data is very
precise; it will vary from head to head due to variances in source intensities, calibration
placement, and manufacturing tolerance. Because of this, each optics head contains its own
individual "XML" file.
Any time an optics head or hard drive is replaced, it may be necessary to reload the
associated calibration file.
Note - The ATLAS software Will Not operate if it detects a mismatch between the
calibration file and optics assembly.
Special Tools / Equipment Required
Calibration file (Possible Sources)
o If replacing the optics head, you should receive a new calibration file on
CDROM.
o If reloading a suspected corrupt or damaged file on an existing instrument, you
can find the original file on the data partition of the hard drive.
Service Manual
Confidential and Proprietary
4-4
ATLAS Model 9000
63041A0208