Power Supply (P/S) Removal - Zeiss ATLAS 9000 Service Manual

Corneal topography system
Hide thumbs Also See for ATLAS 9000:
Table of Contents

Advertisement

Parts Removal/Replacement

3.8 Power Supply (P/S) Removal

1) Remove the Operator panel (Section 3.3).
2) Remove the power input cable (Patient side of P/S).
3) Disconnect the distribution harness (P/S to Motherboard).
4) Remove the 2 screws from the distribution harness side.
5) Slide the module toward the front of the instrument to release.
Note – The Motherboard Power Connector Support Tool is required when reconnecting
the power connector to the Motherboard. Place the support tool underneath the
Motherboard at the middle of the power connector to keep the Motherboard from
flexing while attaching the power connector.
Power
Connector
Support Tool
Service Manual
Figure 3.9 Power Supply
Confidential and Proprietary
3-10
ATLAS Model 9000
Motherboard
Power Connector
63041A0208