12
Data Communication
Overview of Incubator Parameters (Addresses 2xxx)

Reading Basic Parameters

Address
2000
2010
2020
2030
204a
204b
*1)
Example of unit status and (error) status control loops
(for details see error messages)
*2)
All values have two decimal places.
S
i160 LK / i250 LK
TERI-CYCLE
12-6
Description
*1)
(Error-) 
Device status
Status of the control loops
temperature,
CO
, O
, rH, ref. temp.
2
2
Specified, actual and reference
*2
temperature
)
Specified and actual CO
*2)
content
Specified and actual O
content
2
Actual water level (100% or 0%)
Low humidity indicator (1 active, 0
inactive)
Remarks
33 bytes / hexadecimal value
with the format
xxxxxxxx;xxxx; ...
;xxxx;xxxx;xxxx
23 bytes / decimal value
with the format 
+xxx.xx;+xxx.xx;+xxx.xx
15 bytes / decimal value
2
with the format +xxx.xx;+xxx.xx
*2)
15 bytes / decimal value
with the format +xxx.xx;+xxx.xx
7 bytes / decimal value
with the format +xxx.xx
2 bytes / hexadecimal
with the format xx