Thermo Scientific STERI-CYCLE i160 LK User Manual page 118

Co2 incubator
Table of Contents

Advertisement

6
Handling and Control
User Configuration
Logging cycle
10 s
30 s
60 s
120 s
180 s
3600 s
The setting controls the logging cycle in sections of seconds during which the control loop measured
values are logged during the operation of the device and displayed by the trend display
Scaling"
The settings can be made within the value range of 10 seconds to 3600 seconds.
1. To increase the value:
• Press the + key.
2. To reduce the value:
• Press the - key.
3. The value change appears in the display. The New message indicates that the selection has been
changed, but not saved yet.
4. To accept and save the change:
• Press the Enter key.
• The system returns to the Event logging menu.
Note Event logger interval time: 
The logging cycle time does not affect the entries of the error table.
Displaying the Error Table
The error table lists the errors detected by the device-integral monitoring system in descending
chronological order. The most recently detected error is listed at the top position of 22 possible entries.
An entry consists of the control loop in which the error occurred, the date, the time, and an error
description. The error table can be displayed but not edited. If the event display consists of two pages,
the user can browse through the list. The status indicator 001/002 or 002/002 indicates which page of
the two pages is currently being displayed.
Displaying the Error Table
1. Press the Menu key.
2. Select the menu command Event logging.
• The selection dialog shown in
3. Select the menu command Event display.
• The list shown in
S
i160 LK / i250 LK
TERI-CYCLE
6-30
Maximum period displayed
22.5 hours
2.8 days
5.6 days
11.2 days
16.8 days
336 days
on
page
6-43).
Figure 6-34
Figure 6-31
is displayed.
appears on the display.
("Trend Display
Thermo Scientific