6
Handling and Control
User Configuration
Reminder Interval Setting
The reminder intervals are integral components of the alarm and monitoring system of the device
control. For the two essential functions steri-run and auto-start as well as for routine service work, the
user can set dates that trigger an alarm whenever they occur. 
The counting begins at 00:00 hrs of the day on which the previously set reminder interval has elapsed.
On the due date, the display shows a reminder message for the activated reminder interval:
• Steri-run: Please run Steri-Run routine.
• auto-start: Please run auto-start routine. Appears upon successful completion of the Steri-Run
decontamination routine.
• Service interval: Request service. The service message can be confirmed. In this case, the Request
Service icon is displayed.
After the routines have been run successfully, the reminder messages are hidden.
Factory Settings
Steri-run decontamination routine
auto-start routine
Service interval
HEPA interval
Reminder Interval Setting
1. Press the Menu key.
2. Select the Reminder intervals menu option.
• The selection dialog shown in

Figure 6-29. Choose the Desired Reminder Interval Function.

3. Select the appropriate menu option, for example Steri-run.
• The input dialog shown in
S
i160 LK / i250 LK
TERI-CYCLE
6-26
90 days
Off
Off
365 days
Figure 6-29
is displayed.
Figure 6-29
is displayed.