The data sets can be stored in the metaformat *.CSV.
1. Read the data sets:
• Press the READ key.
2. The progress of the data transmission is indicated in the three text boxes:
• AMOUNT DATA: Total number of transmitted data sets.
• AMOUNT TELEGRAMS: Therein the number of transmitted event entries.
• AMOUNT ERRORS: Therein the number of transmitted error messages.
Note Duration of the data transmission:
As the data logger can contain up to 10,000 data sets, the data transmission to a PC may take some
time.
SERVICEFILE
The submenu SERVICEFILE is used for reading error informations and for creating a service file from
it, saved with the proprietary extension *srf. The service file is transmitted to the Technical Service of
Thermo Fisher Scientific for fault analysis.
1. Create the service file:
• Press the CREATE SERVICEFILE key.
• Define file name and saving directory in the Windows dialog box.
2. Start saving process:
• Press the OK key.
Note Duration of the procedure:
The compilation of device informations and the creation of the service file may take some time.
Thermo Scientific
Program STERI-CYCLE i160 LK / STERI-CYCLE i250 LK
S
TERI-CYCLE
12
Data Communication
i160 LK / i250 LK
12-27