Thermo Scientific STERI-CYCLE i160 LK User Manual page 170

Co2 incubator
Table of Contents

Advertisement

11
Specifications
STERI-CYCLE i160 LK
S
i160 LK
TERI-CYCLE
Name
Mechanical
Outer dimensions (W x H x D)
Inner dimensions (W x H x D)
Inner space volume,
of which useable volume
Inserts (B x T)
Quantity of components delivered
Quantity, maximum
Surface loading capacity, maximum
Overall device loading capacity, maximum
Weight
Weight, without accessories
Thermal
Thermal safety features to DIN 12880:2007-05
Ambient temperature range
Ambient temperature for stacked devices
Temperature control range, incubation operation
Temperature deviation from set value, over time (DIN
12880, part 2) at 37 °C
Temperature deviation from set value, spatial (DIN
12880, part 2) at 37 °C
Duration of the auto-start routine: to 37 °C
Ambient temperature 20 °C
Heat transfer to environment:
at 37 °C
during steri-run decontamination
Humidity
Water qualities
Capacities:
Incubation operation
S
i160 LK / i250 LK
TERI-CYCLE
11-2
*1
)
Unit
Value
mm
637 x 905 x 790
(25.08 x 35.63 x 31.10 inch)
mm
470 x 607 x 576
(18.50 x 23.90 x 22.68 inch)
l
approx. 165
l
approx. 100
mm
423 x 465 (16.65 x 18.31 inch)
pcs.
3
pcs.
11
kg
10 per insert
kg
30
kg
83
kg
78.4
Class 3.1 
(Overtemperature controller,
providing alarming function
upon excessive temperature
condition)
°C
+18...34
°C
+18...28
°C
RT + 3...55
°C
± 0.1
°C
<= ± 0.3
h
5...10
kWh/h
0.06
kWh/h
0.59
Electrical resistance:
50 kOhmcm to 1 MOhmcm
Conductivity: 
1 to 20 S/cm
l
max. 3 / min 0.5
Thermo Scientific

Advertisement

Table of Contents
loading

This manual is also suitable for:

Steri-cycle i250 lk

Table of Contents