SQA-V
and
Service Manual 24_Apr_2014
After cleaning:
Press SERVICE > ESC. MAIN MENU will be displayed.
Select TEST NEW PATIENT and re-run test.
If this message is displayed again, reboot the device:
Turn the system OFF then back ON at the main switch on the rear panel.
Press the front panel ON/OFF key to begin Auto-Calibration and Stabilization.
From MAIN menu: Select TEST NEW PATIENT and rerun.
Call technical support if this message is displayed again. Prepare for technical
support by printing a copy of the SQA-V internal parameters.
4. Concentration Out of Range
Normal Testing:
A message will appear indicating that the tests results for SPERM CONC. and/or MSC
are beyond the dynamic range established by the manufacturer for testing. This
message will appear if the SQA-V reads:
SPERM CONC. > 500 M/ml or
MSC > 450 M/ml
Review sample handling technique (see Appendix "Filling the SQA-V Capillary").
Re-test the sample in a new SQA-V capillary. If the message appears again, reboot
the system.
Call for technical assistance if problem persists.
Control Testing:
A message will appear indicating that the CONTROL is out of the dynamic range for
control testing (CONCENTRATION > 118 M/ml) due to:
Dehydrated control media
Bubbles in the testing capillary
Defective SQA-V testing capillary
Check control media expiration date.
Run new control in a new SQA-V testing capillary.
Call for technical assistance if problem persists.
33