Configuration=>Diagnostic - Seitron CHEMIST 600 BE GREEN Use And Maintenance

Combustion analyzer
Table of Contents

Advertisement

11.7
Configuration→Diagnostic
04/03/16
10:00
Diagnostic
Sensors
Pump
On site cal.
Gas probe
Hardware
PARAMETER
Displays information on the state and calibration of the electrochemical sensors:
Ok
absent
err data
unknown
err pos
err cal
err curr
Sensors
err cfg
Also, from this screen the user can access the identification data of the sensor: type, serial
number, date of manufacture and calibration. There are also the measured currents; in this way
it is possible to perform a quick diagnosis in the event of a malfunction.
SEE SECTION
In this submenu the user can temporarily turn the gas suction pump on or off. Also, it is possible
to view the actual flow rate of the pump in liters per minute. It will not be possible to turn off the
pump during an autozero cycle.
Pump
SEE SECTION
It is possible to make a recalibration of the instrument's gas sensors with suitable known
concentration gas cylinders. For the sensors sensible to other gases, called interfering gases
(for example NH
relevant interfering gases.
The sensor recalibration procedure is protected by password: ask Seitron Assistance
On site cal.
center.
SEE SECTION
Tests the tightness of the gas probe pneumatic path.
SEE SECTION
Gas probe
At instrument turn on the firmware performs a full check on the physical efficiency of all types of
HW memories installed on the instrument, as well as on the integrity of the data stored into
them. Any issue is displayed in the screen 'Memories Diagnostics'. Should this happen it is
advisable to turn the instrument off and then on again. In case the problem is permanent or
frequently recurring, the user should contact the Service Center reporting the error code shown
Hardware
by the instrument.
SEE SECTION
KEY
ESC
CONTEXT KEY
OK
No problem detected
The sensor was not detected
Memory data error of the sensor
It is necessary to update the FW of the device
The sensor has been installed in the wrong position
Calibration error (sensor not calibrated)
Currents outside the range
Do not use this sensor as it has not been accepted on the screen "types of sensors".
11.7.1.
11.7.2.
, H
, H
S, SO
, ...) it is possible to perform the on site calibration also for the
3
2
2
2
11.7.3.
11.7.4.
11.7.5.
70
Activate the context keys shown on the
display.
Returns to the previous screen.
Selects the available parameters.
Enters in the selected parameter setting.
Selects the available parameters.
DESCRIPTION
FUNCTION
FUNCTION