15.5 Expandability to 6 sensors

The CHEMIST 600 BE GREEN Emissions Analyzer can be expanded up to 6 cells.
The upgrading of the number of sensors can be easily done by the user by performing the following directions:
- The expandable instruments are arranged in a way to accept up to a maximum of 6 cells.
- Identify, with the help of
added to the existing configuration (Seitron delivers all FLEX-series sensors already pre-calibrated and ready
to use).
- To install the new sensors follow all the steps described in the paragraph
replacement'.
THE INSTRUMENT AUTOMATICALLY DETECTS WHEN AN ADDITIONAL SENSOR IS
INSTALLED OR HAS BEEN REMOVED. THE SCREEN 'SENSORS CONFIGURATION'
ALLOWS TO ACCEPT THE NEW PROPOSED CONFIGURATION OR TO IGNORE THE
CHANGE DETECTED.
IN THIS SCREEN ARE SHOWN, FOR EACH POSITION, THE FOLLOWING MESSAGES:
EXAMPLE OF AN 'NO' SENSOR IN POSITION 3 REPLACED WITH AN 'NO2' SENSOR:
NO→NO
A SENSOR DIFFERENT FROM THE PREVIOUS ONE HAS BEEN DETECTED.
2
EXAMPLE OF A NEW SENSOR INSTALLED IN POSITION 4 (PREVIOUSLY NOT PRESENT):
SO
→□
A NEW SENSOR HAS BEEN DETECTED.
2
paragraph 15.2 'Sensor types and relevant positioning'
'MAINTENANCE'
162
the sensor(s) which must be
under 'gas sensors