10.7 END OF THE ANALYSIS

- At the end of the combustion analysis, carefully remove the sample probe and remote air temperature probe, if
used, from their relative stacks, taking care not to get burnt.
- It is recommended to purge the analyzer with clean fresh air for at least 5 to 10 minutes before turning off the
instrument by pressing the On/Off key.
At this point, if the instrument has detected a high concentration of CO and/or NO, a self-cleaning cycle will be
initiated during which the pump will draw fresh outside air until the gas levels drop below acceptable values.
At the end of the cycle (lasting no longer than 3 min.) the instrument will switch itself off automatically.
Note: It is always suggested to make the instrument purge with fresh clean air for at least 5 to 10 minutes
before switching it off.
WHEN THE SMOKE PROBE IS PULLED OUT OF THE CHIMNEY, CONDENSATE MIGHT
FORM INSIDE THE PROBE PIPE AND IN HTE ANTI-CONDENSATION TRAP.
WE RECOMMEND TO ACCURATELY CLEAN ALL THE PARTS BEFORE PUTTING BACK
THE PROBE AND THE ANTI-CONDENSATION TRAP IN THE SUITCASE.
IN ORDER NOT TO DAMAGE THE INSIDE PART OF THE SUITCASE, MAKE SURE THAT
THE PROBE METALLIC ROD IS COLDER THAN 60°C.
Cleaning of the smoke probe
- When you finish using the sample probe clean it thoroughly as described below before returning it to its case:
- Disconnect the sample probe from the instrument and from the water trap (Fig. a-b) then blow a jet of clean air
into the hose of the probe (refer to Fig. b) to remove any residual condensate that may have formed within.
Fig. a
Cleaning hose
40
Fig. a
Fig. b
K400000000SE 041891 010823