Maintenance; Routine Maintenance; Preventive Maintenance - Seitron CHEMIST 600 BE GREEN Use And Maintenance

Combustion analyzer
Table of Contents

Advertisement

16.0 MAINTENANCE

16.1 Routine maintenance

This instrument was designed and manufactured using top-quality components. Proper and systematic
maintenance will prevent the onset of malfunctions and will increase instrument life altogether.
General guidelines for operations of the analyzer include the following:
Do not expose the instrument to substantial thermal shocks before use. If this happens, wait for the
temperature to return to normal working values.
Do not extract flue gas samples directly without using a particulate/water trap.
Do not exceed sensor overload thresholds.
When the analysis is completed, disconnect the water trap and hoses and let the analyzer purge with clean
fresh air for at least 5 to 10 minutes, or at least until the displayed parameters return to their original values in
air.
Clean the filter unit when necessary, replacing the particulate filter and blowing air through the sample probe
hose to eliminate any condensate that may have formed.
Do not clean the instrument with abrasive cleaners, thinners or other similar detergents.

16.2 Preventive maintenance

At least once a year send the instrument to a SERVICE CENTER for a recalibration of the analyzer and
maintenance check-up SEITRON's highly qualified staff is always at your disposal and will provide you with all
the sales, technical, application and maintenance details required.
The service center will always return the instrument to you as new and in the shortest time possible. Calibration is
performed using gases and instruments comparable with National and International Specimens. Annual servicing
is accompanied by a specific calibration certificate that is a guarantee of perfect instrument performance as
required by UNI 10389-1, besides being indispensable for users wishing to maintain ISO 9000 status.
165
K400000000SE 041891 010823