Power On - Off; Starting The Device - Seitron CHEMIST 600 BE GREEN Use And Maintenance

Combustion analyzer
Table of Contents

Advertisement

9.0

POWER ON - OFF

9.1

Starting the device

Press and hold for a
few seconds.
DURING INSTRUMENT AUTOZERO, THE SAMPLING PROBE MUST NOT BE INSERTED
IN THE CHIMNEY.
During autozero, you can only use the menus that do not require autozero.
!
ERROR
Autozero failed.
F1: Autozero
F2: Analysis
F3: Diagnostic
04/03/16
Combustion analysis
O
2
%
CO
!
WARNING
λ,n
Do not insert the gas
probe in the chimney
T
T
Press any key
ΔT
°C
Qs
%
ηs
%
This error message is displayed if the autozero of the device is not
Repeat?
successfully completed.
10:00
Combustion analysis
O
4.2
2
%
CO
9.3
2
%
λ,n
1.2
T flue
190
°C
T air
15.
°C
ΔT
74.
°C
Qs
8.6
%
ηs
91.4
%
27
04/03/16
10:00
Combustion analysis
O
4.2
2
%
CO
9.3
2
%
λ,n
1.25
T flue
190.1
°C
T air
15.4
°C
ΔT
74.7
°C
Qs
8.6
%
ηs
91.4
%
K400000000SE 041891 010823
04/03/16
10:00
4.2
9.3
1.25
190.1
15.4
74.7
8.6
91.4