Metrohm 942 Extension Module Vario Manual page 45

Hide thumbs Also See for 942 Extension Module Vario:
Table of Contents

Advertisement

■■■■■■■■■■■■■■■■■■■■■■
Accessories
942 Extension Module Vario ONE/Deg (2.942.1060)
3 Reinserting the inlet valve into the pump head
Insert the inlet valve into the inlet valve holder (the seal must be
visible) (1).
Screw the inlet valve holder up into the pump head (2) and
tighten it firmly by hand and then retighten it one additional ¾
turn using the adjustable wrench (3).
Refasten the connection capillary to the connection of the eluent
aspiration tubing (14-7).
Removing the pump head
For this step, you need the following accessories:
4 mm hex key (6.2621.030)
Removing the pump head
Prerequisites:
Is the high-pressure pump switched off?
Has the pressure been released?
Is the instrument switched off?
5 Operation and maintenance
■■■■■■■■
37

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 942 Extension Module Vario and is the answer not in the manual?

Table of Contents