3. Specification

● Applicable standards
 
IEC 61010-1,-2-030 
 
 
Measurement CAT III 600V Pollution degree2
 
 
Measurement CAT IV 300V Pollution degree2
 
IEC 61010-031  Standard for hand-held probes
 
 
MODEL7165A(CAT IV 600V)
 
 
MODEL7264(CAT IV 600V)
 
 
MODEL7265(CAT IV 600V)
 
 
*   W hen KEW3125A/3025A and the test lead 
are  combined  and  used  together,  whichever 
is lower category either of them belongs to is 
applied.
 
IEC 61326-1,-2-2  
 
IEC 60529  IP40
 
EN 50581 
 
● Measuring range and accuracy 
  (Temperature, humidity: 23±5Cº, 45-75%RH)
 【Insulation resistance tester】 :KEW3125A/3025A
Rated voltage
250V
500V
0.0-100.0MΩ
0.0-99.9MΩ
Measuring 
80-1000MΩ
Range
Display 
0.0 ‒ 105.0MΩ
0.0-1050MΩ
range
Open circuit 
DC 250V
DC 500V
Voltage
+10%, -10%
+20%, -10%
Rated 
Current
Short-circuit 
Current
Accuracy
www.calcert.com
15
10
20
5
25
0
30
EMC standard
RoHS Directive
1000V
2500V
0.0-99.9MΩ
0.0-99.9MΩ
80-999MΩ
80-999MΩ
0.80-2.00GΩ
0.80-9.99GΩ
8.0-100.0GΩ
0.0M-2.10GΩ
0.0M-105.0GΩ
DC 1000V
DC 2500V
+20%, -0%
+20%, -0%
1mA or more,
1mA or more,
1.2mA or less
1.2mA or less
(at 1MΩ load)
(at 2.5MΩ load)
1.5mA±0.5mA
±5%rdg±3dgt 