5. Getting started
5-1 Checking the battery voltage
   (1) Set the Range switch to any position other than OFF .
   (2)   W hen the Battery mark shown at the upper left on the LCD 
is last one level 
 , the batteries are almost exhausted. 
          R eplace  the  batteries  with  new  ones  to  perform  further 
measurements.
          T he instrument operates properly even if under such a low 
battery level, and it may not affect on the accuracy.
          W hen  the  empty  Battery  mark 
voltage is below the lower limit of the operating voltage. So 
the accuracy cannot be guaranteed.
          P lease  refer  to  clause  7.  Battery  Replacement,  for  the 
battery replacement procedures.
5-2 Connecting test leads
     I nsert  the  test  lead  firmly  to  the  connector  terminal  on  the 
instrument.  Connect  the  Line  Probe(red)  to  the  Line  terminal, 
the  Earth  Cord(black)  to  the  Earth  terminal  and  the  Guard 
Cord(green)  to  the  Guard  terminal.  A  Guard  cord  connection 
may not be required, if there is no need to establish a guard. 
# DANGER
●   I f  the  Test  button  is  pressed  when  the  Range  switch  is  at 
the insulation measurement positions, high voltage may be 
applied to the test leads and you may get an electric shock.
●   S top using the test lead if the outer jacket is damaged and 
the inner metal or color jacket is exposed.
̶  13  ̶