KYORITSU 4118A Instruction Manual

KYORITSU 4118A Instruction Manual

Digital psc-loop
Hide thumbs Also See for 4118A:

Advertisement

INSTRUCTION MANUAL
DIGITAL PSC-LOOP TESTER
MODEL 4118A
KYORITSU ELECTRICAL INSTRUMENTS
WORKS, LTD.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 4118A and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for KYORITSU 4118A

  • Page 1 INSTRUCTION MANUAL DIGITAL PSC-LOOP TESTER MODEL 4118A KYORITSU ELECTRICAL INSTRUMENTS WORKS, LTD.
  • Page 2: Table Of Contents

     CONTENTS 1. SAFE TESTING .................. 1 2. PROCEDURE OF REMOVING COVER ........... 4 3. FEATURES ..................5  3.1 Instrument Layout ................ 5  3.2 Test Lead ..................6  3.3 Features ..................7 4. SPECIFICATIONS ................9 5. OPERATING INSTRUCTIONS ............10  5.1 Initial Checks ................
  • Page 3: Safe Testing

    1. This instrument must only be used by a competent and trained person and operated in strict accordance with the instructions. KYORITSU will not accept liability for any damage or injury caused by misuse or non-compliance with the instructions or with the safety procedures.
  • Page 4 Failure to follow the instructions may cause injury, instrument damage and/or damage to equipment under test. Kyoritsu is by no means liable for any damage resulting from the instrument in contradiction to this cautionary note. ●   Never open the instrument case - there are dangerous voltages present.
  • Page 5   CAUTION ●   For those testers without the D-LOK circuitry (models 4118A) all RCD's (RCCB, ELCB) in the circuit must be by-passed for the duration of the test (except on loop-2000Ω range).   Do not operate the RCD Test Button with the RCD by-passed.
  • Page 6: Procedure Of Removing Cover

     2. PROCEDURE OF REMOVING COVER Model 4118A have a dedicated cover to protect against an impact from the outside and prevent the operation part, the LCD and the connector socket from becoming dirty. The cover can be detached and put on the back side of the main body during measurement.
  • Page 7: Features

     3. FEATURES 3.1 Instrument Layout ⑤ ① ② Wiring correct (Green LED's) Reversed L/N terminals (Red LED) ③ ④ Connector Fig. 3 1.... LCD display 2.... Wiring check LEDs 3.... Test button 4.... Range switch 5.... Connector   DANGER ●   Use original test lead only. ●...
  • Page 8: Test Lead

    3.2 Test Lead The instrument is supplied with Model 7125 lead at socket outlets and Model 7121B distribution board lead. (1) Model 7125 Mains lead mentioned in this instruction manual is Model 7125(EU) European SCHUKO plug, however, there are some other types for the areas, and the shipping is made along with the suitable plug-shaped cable according to the areas of the globe as follows;...
  • Page 9: Features

    Choose and use the test leads and caps that are suitable for the measurement category. When the instrument and the test lead are combined and used together, whichever lower category either of them belongs to will be applied. Mechanical safety part: Barrier: provides protection against electrical shock and ensuring the minimum required air and creepage distances.
  • Page 10 3.3-3 Model 4118A have the following features: ● Battery is not used All models are not battery-operated, but operate by the voltage supplied from the system. ● Wiring check Three LEDs indicate if the wiring of the circuit under test is correct. The P-E and P-N LEDs illuminate when the wiring polarity of the circuit under test is correct.
  • Page 11: Specifications

     4. SPECIFICATIONS ● Measurement Specification Loop Impedance Nominal test current Range Measuring range Accuracy at 0 Ω external loop 20 Ω 0.00 〜 19.99 Ω 25A / 20ms 200 Ω 0.0 〜 199.9 Ω 2.3A / 40ms ± (2%rdg + 4dgt) 2000 Ω...
  • Page 12: Operating Instructions

    ● Operating Instrumental Uncertainty of Loop Impedance (61557-3) Measuring range to keep Maximum percentage operating Range operating error instrumental uncertainty 20 Ω 0.35 〜 19.99 Ω 200 Ω ± 30% 20.0 〜 199.9 Ω 2000 Ω 200 〜 1999 Ω The influencing variations used for calculating the operating instrumental uncertainty are denoted as follows: Temperature:0℃...
  • Page 13: Measurement Of The Loop Impedance

       Do not use this instrument to check the correct wiring of the power supply. Kyoritsu will not be held liable for any accident that may result from incorrect wiring of the power supply line. (3) Voltage Measurement  ...
  • Page 14: Measurement Of Prospective Short Circuit Current

    20° 30° -13% 5.3 Measurement of Prospective Short Circuit Current   - (Models 4118A)  (1) Set the instrument to the 20kA range.  (2) Connect the test lead to the instrument.  (3) Attach the plug to the socket to be tested.
  • Page 15: Detailed Explanation

     (4) Check that the LED's are lit in the sequence indicated in section 5.1. If not, disconnect from the mains and check the wiring at the socket.  (5) Press the "Press to Test" button. The prospective short circuit current (PSC) will be directly displayed on the LCD with the appropriate units. This will remain for 3s and then revert to AC voltage display.
  • Page 16   In the event of a fault, the Fault loop impedance should be low enough (and the Prospective Fault current higher enough) in order to have the automatic disconnection of supply by the installed protection device within prescribed time interval.  ...
  • Page 17 Ω RA (at 25V)  Note:    ● The loop tester models 4118A measure the fault loop impedance that is a value normally a little bit higher of RA.      But, if the electrical installation is protected considering the loop impedance value, also the RA formula will be fulfilled.
  • Page 18 Practical example of verification of the protection in a TT system according to the international Standard IEC 60364. Fig.8 For this example max value is 1667Ω, the loop tester reads 12.74Ω, it < means that the condition RA 50/Ia is respected. It is fundamental for this example to test also the RCD to ensure that operation takes place quickly enough to respect the safety requirements.
  • Page 19 0.27 0.39 1020 0.22 Using the current ranges on models 4118A can be also tested the Prospective Fault current. Prospective Fault current measured by instruments must be higher than Ia of the protective device concerned Practical example of verification of the protection in a TN system according to the international Standard IEC 60364.
  • Page 20: Measurement Of Old-Tt System

    Max value of Zs for this example is 2.1Ω (16A gG fuse, 0.4s) the loop tester reads 1.14 Ω (or 202 A on Fault current range) it means that the condition Zs < Uo/Ia is respected. In fact the Zs of 1.14Ω is less than 2.1Ω (or the Fault current of 202 A is more than Ia of 110A).
  • Page 21: Measurement Of Line Impedance And Prospective Short Circuit Current

      WARNING (OLD-TT system only) ●   DO NOT PRESS the "Test button" if the display reads a value of 220V! 6.3 Measurement of Line Impedance and Prospective Short Circuit Current Line Impedance on single-phase system is the impedance measured between phase and neutral terminals.
  • Page 22: Servicing

      WARNING ●   This instrument is intended only for use in single phase operation at 230V +10% -15% AC phase to earth or for use in OLD-TT system phase to neutral. ●   If the overheat symbol appears in the display (   ) disconnect the instrument from the mains supply and allow to cool down.
  • Page 24 DISTRIBUTOR 92-2233 15-02 Kyoritsu reserves the rights to change specifications or designs described in this manual without notice and without obligations.

Table of Contents