Scheduled maintenance
3. Check the expiration date on the pads.
4. Listen for the voice prompts.
5. Close the lid and observe the change of the STATUS
INDICATOR to RED. After approximately 5 seconds, verify that
the STATUS INDICATOR returns to GREEN.

Annual maintenance

Perform the following tests annually to confirm that the diagnostics are
functioning properly and to verify the integrity of the case.
Check the integrity of the pads and circuitry:
1. Open the AED lid.
2. Remove the pads.
3. Close the lid.
4. Confirm that the STATUS INDICATOR turns RED.
5. Open the lid and confirm that the Pad indicator is lit.
6. Reconnect the pads and close the lid.
7. Make sure the expiration date is visible through the clear window
of the lid.
8. Check to make sure that the STATUS INDICATOR is GREEN.
If the pads are not installed properly, the PAD indicator will
illuminate. Contact Cardiac Science Technical Support (see
Contact information
Cardiac Science representative.
9. Open the lid and confirm that no diagnostic indicators are lit.
10. Check the expiration date of the pads; if expired, replace them.
11. Check the pads packaging integrity.
12. Close the lid
Check the Integrity of the Service Indicator (LED) and Circuitry:
1. Immediately after opening the AED lid, press and hold the Shock
button and confirm that the Service LED is lit (for the Powerheart
G3 model 9390E only).
2. Release the Shock button (for the Powerheart G3 model 9390E
only).
3. Close the lid.
4. Verify that the STATUS INDICATOR remains RED.
Powerheart® AED G3 PLUS 9390A AND 9390E
on page 1-2) or outside the U.S., your local
70-02062-02 A
5-5