AED indicators

Diagnostic panel

The diagnostic panel has the following indicators:
1
2
3
4
Smartgauge battery status indicator
0%
from right to left, as battery capacity decreases. When the green LEDs
go out and the red LED lights up, replace the battery.
Note: When the red LED initially lights up—upon lid opening or at
any time during a rescue—a BATTERY LOW prompt will be issued at
once. However, the AED is capable of delivering at least 9
defibrillation shocks after the first BATTERY LOW prompt is issued.
When the AED battery cannot deliver any more shocks, the AED
shows BATTERY LOW on the text display, and the red battery LED
illuminates. To continue the rescue, leave the lid open, remove the
battery, and replace with a fresh battery. If battery replacement takes
longer than 60 seconds, the first rescue will be terminated and a second
rescue will begin upon insertion of battery.
Note: When the battery is depleted, neither the LED nor the text
display illuminates.
Powerheart® AED G3 PLUS 9390A AND 9390E
2 3
0%
100%
1
Smartgauge™ battery indicator
Pads indicator
Service indicator
Shock Button (Powerheart G3 model 9390E only)
100%
4
The Smartgauge Battery Status
Indicator has five LEDs, four green and
one red. The right four green LEDs
display the remaining capacity of the
battery much like a fuel gauge. With
use, the green LEDs gradually go out,
70-02062-02 A
3-3

Advertisement