Table 3-8: Other prompts (continued)
Voice prompt
Text display
"Open lid to continue
OPEN LID
rescue. "
TO CONTINUE RESCUE
"Rhythm changed. Shock
RHYTHM CHANGED
cancelled. "
SHOCK CANCELLED
"Remove cable to continue
REMOVE CABLE
rescue. "
TO CONTINUE RESCUE
"Communications mode"
COMMUNICATIONS MODE
"Service required"
SERVICE REQUIRED
Powerheart® AED G3 PLUS 9390A AND 9390E
Situation
When the lid is inadvertently closed during a rescue, this prompt will repeat for
15 seconds.
When the device is prepared to shock then detects a change in rhythm and
therefore cancels the shock.
When a serial communication cable is connected to the AED during a rescue, the
phrase repeats until the cable is disconnected.
When the lid is open and the serial communication cable is plugged into the
AED.
Occurs after the self-tests determine that the AED is not functioning properly.
The prompt "SERVICE REQUIRED" will be heard when the lid is opened. The red
SERVICE indicator will illuminate. After closing the lid, an alarm beep will be
heard until the battery is removed or becomes completely depleted.
RescueCoach™ voice prompts and text display
70-02062-02 A
3-13