Technical Data
Table 6-1: Parameters (continued)
Parameter
Dimensions
Weight (Batteries and Pads) 9390: 3.10 kg (6.6 lb)
Environmental Operation
and Standby Conditions
Shipment and Transport
environmental Conditions
(for up to 1 week)
Pads
9146 Lithium Battery
Specifications
Batteries and Capacitor
Charge Times
70-02062-02 A
6-2
Detail
Height: 8 cm (3.3 in)
Width: 27 cm (10.6 in)
Depth: 31 cm (12.4 in)
Temperature: 0°C to 50°C (32°F to 122°F)
Humidity: 5% to 95% (non-condensing)
Pressure: 57kPa (+15,000ft) to 103kPa (-500ft)
Temperature: -30°C to 65°C (-22°F to 149°F)
Humidity: 5% to 95% (non-condensing)
Pressure: 57kPa (+15,000ft) to 103kPa (-500ft)
Self-adhesive, disposable defibrillation pads
Minimum combined surface area: 228cm
length of lead wire: 1.3m
Output voltage: 12VDC
Batteries are non-rechargeable
Lithium content: 9.2g
Check local regulations for disposal information
Full Operational Replacement Guarantee (from date of
installation): 4 Years
Estimated Shelf Life (from date of manufacture): 5 Years
Typical Shocks: 290 shocks
Note: The battery operating life depends on the type of
battery, device settings, actual usage, and environmental
factors. Battery was tested with G3 Plus device with
Standard prompt set and CPR set to 60 seconds.
A new battery, after the AED has delivered 15 300VE
shocks, typically takes 10 seconds to charge the AED to
maximum energy.
A battery with reduced capacity will take longer to
charge the AED.
Powerheart® AED G3 PLUS 9390A AND 9390E
2
Extended