Cardiac Science POWERHEART AED 9390E Operator's And Service Manual page 48

Table of Contents

Advertisement

Getting Started
Table 3-7: Pad issues
Voice prompt
"Make sure pad connector is plugged
into AED. "
"Press pads firmly to patient's bare
skin. "
Table 3-8: Other prompts
Voice prompt
Text display
"Battery low"
BATTERY LOW
"Analysis interrupted. Stop
ANALYSIS INTERRUPTED
patient motion. "
STOP PATIENT MOTION
70-02062-02 A
3-12
Text display
CHECK CONNECTOR
IS PLUGGED INTO AED
PRESS PADS FIRMLY
TO BARE SKIN
Situation
Occurs once when the battery voltage becomes low, although a rescue can
continue for approximately 9 more shocks. When the battery is too low to do a
rescue, the following will occur:
BATTERY LOW shows on the LCD
Smartgauge battery status indicator turns red
AED beeps once every 30 seconds while the lid is closed
You must replace the battery before continuing with the rescue. If completely
depleted, all AED activity will terminate.
When the AED detects ECG noise artefact, stop moving or touching the patient.
Remove other electronic devices within a 5 meter radius.
Situation
Prompts when defibrillation pads connector is not correctly inserted into
pad socket.
Prompts when better pad connectivity to the patient's skin is required
because impedance is too high.
Powerheart® AED G3 PLUS 9390A AND 9390E