Immunity test
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
61000-4-11
Power frequency (50/
60 Hz) magnetic field
IEC 61000-4-8
U
Note:
T
70-02062-02 A
1-14
IEC 60601 test level
<5% U
T
(>95% dip in U
T
cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
(30% dip in U
T
for 25 cycles
<5% U
(>95% dip in U
T
for 5 sec.
3 A/m
is the a.c. mains voltage prior to application of the test level.
Compliance level
Not applicable
) for 0.5
)
T
)
T
80 A/m
Electromagnetic environment—
guidance
Power frequency magnetic fields should
be at levels no higher than those
characteristic of a typical location in
typical heavy industrial and power
plants and the control rooms of H.V. sub-
stations.
Powerheart® AED G3 PLUS 9390A AND 9390E