Table 3-2: Preparation (continued)
Voice prompt
"When patient's chest and torso are exposed, remove
square foil package from lid of AED."
"Tear open foil package across dotted line and remove
pads. "
"Next, separate one of the white pads completely from
blue plastic liner. Begin peeling from the tabbed corner. "
"Firmly place the pad without the liner on the patient,
exactly as illustrated. This pad can be placed on either of
the two locations shown. "
"Next, peel the blue plastic liner off of the second white
pad. "
"Firmly place the second pad on the opposite location,
exactly as illustrated. "
Powerheart® AED G3 PLUS 9390A AND 9390E
Text display
WHEN CHEST IS BARE
REMOVE FOIL PACKAGE
TEAR OPEN PACKAGE
REMOVE PADS
PEEL ONE PAD FROM
BLUE PLASTIC LINER
PRESS PAD FIRMLY
TO CHEST AS SHOWN
PEEL SECOND PAD OFF
BLUE PLASTIC LINER
PRESS PAD FIRMLY
AS SHOWN
RescueCoach™ voice prompts and text display
Situation
Prompts the rescuer to remove the pads from AED lid.
Prompts the rescuer to open the pad package and
remove pads.
Repeats every 3 seconds until the pads are separated.
If a pad has been peeled before the prompt starts,
this prompt will be skipped. This prompt will be
interrupted when pad is peeled.
Prompts the rescuer to place one pad on the patient.
Prompts the rescuer to remove the liner from the
second pad.
Repeats until second pad placement is sensed. If the
pad is placed before prompt starts then this prompt
will be skipped. This prompt will be interrupted when
second pad is placed.