2.6- Install accessories

Continue to keep the device upside down and disconnected from the power grid!
EH-142-US
1
Air hose – filter pre-installed
USE HARD FORCE
EG-110-US
2
Water hose – filter pre-installed
Power cord into socket
3
(Fuse holder in the socket)
EK-410
4
Wired pedal
ONLY IF APLICABLE
EM-145
5
AIRFLOW
USE FORCE

2.7- Check the cord connections

The handpiece cord is not fully connected.
To disconnect the handpiece cord system, unlock the connection and pull at the same time.

2.8- Fix the device

Fix the device to the recommended STATION
9
Follow the quick guide delivered with your STATION
9
Fix the device to the recommended STATION
9
You will find a "Master Screw" provided on the bottom center of the device.
Unscrew the Master Screw first and use it to secure the device firmly to a table or onto the AL-125 device support in your cabinet
(the AL-125 part is available through our after-sales support and dealers).
Fix your device with the provided "Master Screw" in order to ensure that the unit cannot be removed without the use of a
tool.
Check the position of the medical device so that it corresponds to your line of sight and the characteristics of your personal
workstation (the lighting and the distance between the user and the device). The device must remain quickly and easily
accessible at all times.
Check that the water and air lines and the power cord do not hinder physical movement.
9 AIRFLOW
STATION or STATION+
®
8
handpiece cord + lock actuator
®
USE FORCE to lock in.
The system is well connected & locked.
Master Screw placement

2.9- Power your device

You can now connect the power cord to the mains grid.
Protective earth is required! Be sure your power grid has an efficient protective earth.
Settings: Voltage: 100-240 Vac - Frequency: 50 to 60 Hz - Operating current: 4 A max.
Scan
for
support

2.10- Installation of the wireless pedal

Insert two (2) AA 1.5V lithium batteries into the wireless pedal. Close the cover and operate your
device. USE ONLY LITHIUM BATTERIES. Close the cover and operate your device.
Risk of fire: only use batteries that have current limiter/short-circuit and over-temperature
protection (compliant to IEC 60086-4:2014 Safety of lithium batteries).
The wireless pedal supplied with your device is already paired and ready to use (Note: A pedal can only command one single
device at a time. Pairing is maintained even if the batteries are removed).
When you receive your new machine, all you need to do is insert the two (2) AA lithium batteries into the wireless pedal and your
device is ready to work.
In case you replace your pedal, you will need to pair it with your device. For instructions, please read the specific Maintenance
& Troubleshooting chapter.
The Bluetooth
radio is automatically disabled (powered off) when a wired pedal is connected.
®
The wireless pedal uses a low power, 8 dBm EIRP max, Bluetooth
Interference may occur in the vicinity of this equipment.
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used
no closer than 30 cm (12 inches) to any part of the device, including cables. Otherwise, degradation of the performance of this
equipment could result.

III- POWDER CHAMBERS

Clinical risk: Only use PLUS Powder with the PLUS Powder chamber.
Clinical risk: Only use PLUS Powder chamber (red) for subgingival treatments.
PLUS
The PLUS Powder chamber is designed for the PLUS Powder. It can be used for supragingival and
subgingival treatments.
Pressure is automatically reduced for compatibility with subgingival treatments, including PERIOFLOW
treatments (Supra applications also possible).
Compatible EMS Powders: PLUS (refer to paragraph "Compatibility" for details).
CLASSIC
The CLASSIC Powder chamber is designed for the CLASSIC Powder and can only be used for
supragingival treatments.
Sodium Bicarbonate: Only use this powder and chamber for supragingival applications.
Compatible EMS Powders: CLASSIC (refer to paragraph "Compatibility" for details).
Check bottle and powder chamber integrity: There should be no crack on the body.
The powder chamber is pressurized during use. Replace faulty parts immediately.
Make sure that the powder chambers are dry.
Use only PLUS Powder for restorations, crowns, bridgework, implants and orthodontics.
Regularly clean the powder chambers. Take care that all parts are fully dry before reuse.
DO NOT sterilize the powder chambers and their caps/parts by steaming or dry thermal reprocessing.
By hand only: remove the powder chamber cap to refill powder up to the indicated MAX level, then
insert the cap back fully onto the bottle.
Pour the powder in freely. The central tube can be fully filled without problem.
DO NOT fill the chamber higher than the indicated MAX level. The powder level will go down slightly
a few minutes after the filling (powder compaction).
Before pressurizing, position the powder chamber into the device. Magnetic attraction will position it
correctly.
DO NOT insert upside-down.
2.4 GHz radio, to communicate with the control unit.
®
®
9