1.1- AIRFLOW
®
pressure setting
Both the PLUS and CLASSIC powder chambers have an integrated dynamic pressure
regulator that automatically set the optimal pressure range for the selected powder chamber
and related powder type as detailed in chapter "Powder Chambers".
The following table shows the static and approximate dynamic pressures
setting:
Pressure settings
0
1
Static (Bar)
/
2.5
CLASSIC dynamic (Bar)
/
1.9
PLUS dynamic (Bar)
/
1.5
1.2- AIRFLOW
BOOST
®
Pressing hard on the center of the wireless pedal activates the BOOST mode and results in an increase of power, as the
following table shows:
Power settings
0
AIRFLOW
®
0
corresponding Boost level

1.3- Wireless pedal battery saving

Each time the wireless pedal is released, it enters into a low power mode. Even if unused for long, it is not required to remove
the batteries.
To avoid an involuntary depletion of the wireless pedal batteries, in case the pedal remains pressed without interruption for 10
minutes, it will automatically enter into switch-off mode.
To resume from the switch-off mode, it is required to first release the wireless pedal and then power cycle the device (switch
off for 30s and then power on again).
Scan
for
support

1.4- Water temperature and acoustic feedback settings

AIRFLOW
®
water temperature is 40°C by default.
To adjust the water temperature or the acoustic feedback, follow the procedure below:
1.
Turn the device ON.
2.
Securely place the AIRFLOW
handpiece back into its holder.
®
3.
Press 0 + 10 simultaneously to access the menu. (See image below – place fingers in the groove below the numbers)
4.
Colors will appear on the numbers:
• 0 to 4 for setting water temperature (5 is not used)
• 6 to 10 for setting acoustic feedback (5 is not used)
Water temperature
11
0
1
2
3
No
25°
30°
35°
Heating
5.
Change the settings according to your wish.
6.
Press the ON/OFF button to save the setting and exit.
Note:
• After a few seconds of keyboard inactivity, the device automatically exits the mode.
10 Dynamic pressures depend on handpiece and powder type too. The listed pressures are for information purpose and referring to the commonly used EL-308
AIRFLOW
Handpiece with DV-082 and DV-048 powders.
®
11 The target temperature is determined into the device's body.
On AIRFLOW
®
side, water temperature decreases along the cord. Air spray also decreases the temperature. Final temperature of AIRFLOW
than 40°C.
12
as per selected powder chamber and user power
10
2
3
4
5
6
7
8
9
2.7
3.0
3.2
3.5
3.7
4.0
4.2
4.5
2.1
2.3
2.6
2.8
3.0
3.2
3.5
3.7
1.7
1.9
2.0
2.2
2.4
2.6
2.7
2.9
1
2
3
4
5
6
7
8
6
7
8
8
8
9
10
10
10
Acoustic feedback
4
6
7
8
9
No
Low
Medium
High
40°
sound
volume
volume
volume
®

II- TREATMENT SEQUENCE

Consult the Treatment Recommendations (FB-648/US) before starting any treatment to the patient.

2.1- Patient and dental professional precautions

For your protection and the protection of your patient, start by rinsing with an antimicrobial mouth rinse.
10
4.7
3.9
It is also recommended to remove the patient's glasses and optical lenses.
3.1
OptraGate for lip and cheek retractor protection is recommended for maximum
patient comfort and improved access to the treatment field. If a retractor is not used,
place cocoa butter or equivalent on the patient's lips.
If PERIOFLOW
clinical probing depth.
Protect yourself with the following measures:
9
10
Wear protective mask
10
Additional personal protective equipment can be used.
High vacuum suction (HVS) and saliva ejector are mandatory. They evacuate the air/powder mixture deviated by the treated
tooth. Study
2.2- AIRFLOW
Angle
Aerosol management:
Use high-speed suction cannula. Positioning the HVS in opposite nozzle direction.
  Direct the jet projections towards the cannula.
Optimal efficiency
  Avoid using the AIRFLOW
  Continuously adapt the angle while working
  Maximal range of usage is between 15°-80°
Distance
  General rule: With AIRFLOW
  Keep the handpiece at 2 to 5 mm during work
  In case of heavy stains, keep the handpiece at max 2mm
  More than 5mm, it will lose efficiency
Movement
  Make continuous semi-circular movement
  Small smileys mesial to distal
  Never hold the handpiece stationary
10
Maximum
volume
12 Aerosols in Dentistry: The Bacterial Contamination of the Room Air During an AIRFLOW
spray is lukewarm, lower
Adrian Lussi. www.ZM.ONLINE.de 12/2020

2.1.1- Patient preparation

Eye protection is mandatory.
treatment has to be performed, radiographs is mandatory to correlate
®

2.1.2- Dental professional preparation

Wear protective glasses
on aerosol management shows that if AIRFLOW
is used as recommended, the risk is negligeable for the clinician.
12
®
and PERIOFLOW
®
®
2.2.1- AIRFLOW
treatment
®

2.2.1.1- Recommended position and movement

Handpiece at 90°
®
MAX, work closer!
®
EMS Nyon
Wash your hands
Wear protective gloves
15° to 80°
2-5mm
Treatment. Marcel Donnet, Klaus-Dieter Bastendorf, Magda Mensi,
®
13