Spectrum Analyzer Measurements
Spectrum Emission Mask Measurement Basic Setup
1.
Select MEASURE > MEASUREMENT > SEM from the right side menu.
2.
Select SETUP and select upper or lower mask (5GNR P>38 dBm or 5GNR P<38 dBm).
3.
Enter the reference channel bandwidth.
4.
Toggle AUTO MAX PWR or set MANUAL MAX POWER.
5.
The SEM measurement results are displayed in a table with pass/fail results.
Spectrum Analyzer MG
PN: 10580-00447 Rev. E
3-16 Spectrum Emission Mask
3-73