Anritsu Field Master Pro MS2090A User Manual

Anritsu Field Master Pro MS2090A User Manual

High-performance handheld spectrum analyzer
Hide thumbs Also See for Field Master Pro MS2090A:
Table of Contents

Advertisement

Quick Links

User Guide
Field Master Pro™ MS2090A
High-Performance Handheld Spectrum Analyzer
MS2090A-0709 9 kHz to 9 GHz (Option 709)
MS2090A-0714 9 kHz to 14 GHz (Option 714)
MS2090A-0720 9 kHz to 20 GHz (Option 720)
MS2090A-0726 9 kHz to 26.5 GHz (Option 726)
MS2090A-0732 9 kHz to 32 GHz (Option 732)
MS2090A-0743 9 kHz to 43.5 GHz (Option 743)
MS2090A-0754 9 kHz to 54 GHz (Option 754)
Anritsu Company
Part Number: 10580-00444
490 Jarvis Drive
Revision: U
Morgan Hill, CA 95037-2809
Published: April 2023
USA
Copyright 2023 Anritsu Company

Advertisement

Table of Contents
loading

Summary of Contents for Anritsu Field Master Pro MS2090A

  • Page 1 MS2090A-0732 9 kHz to 32 GHz (Option 732) MS2090A-0743 9 kHz to 43.5 GHz (Option 743) MS2090A-0754 9 kHz to 54 GHz (Option 754) Anritsu Company Part Number: 10580-00444 490 Jarvis Drive Revision: U Morgan Hill, CA 95037-2809 Published: April 2023 Copyright 2023 Anritsu Company...
  • Page 2 Anritsu Company. Export Management The Anritsu products identified herein and their respective manuals may require an Export License or approval by the government of the product country of origin for re-export from your country. Before you export these products or any of their manuals, please contact Anritsu Company to confirm whether or not these items are export-controlled.
  • Page 3: Table Of Contents

    Contacting Anritsu for Sales and Service ........
  • Page 4 Table of Contents (Continued) 2-10 Settings Menu ............. 2-23 Display Settings .
  • Page 5 Table of Contents (Continued) Appendix B—Upgrading Software Options Introduction ..............B-1 Exporting a Software Configuration File .
  • Page 6 Contents-4 PN: 10580-00444 Rev. U MS2090A UG...
  • Page 7: Chapter 1-General Information

    Software”). Before You Begin • Read the Field Master Pro MS2090A Information, Compliance, and Safety Guide (PN: 10100-00069) for important safety, legal, and regulatory notices before operating the equipment. • Charge the instrument battery using the supplied battery charger until fully charged.
  • Page 8: Additional Documentation

    Field Master Series Cable and Antenna Analyzer Measurement Guide (Option 331) For additional information and literature covering your product, visit the product page of your instrument and select the Library tab: http://www.anritsu.com/en-US/test-measurement/products/ms2090a Document Conventions The following conventions are used throughout the MS2090A documentation set.
  • Page 9: Instrument Description

    General Information 1-2 Instrument Description Instrument Description The MS2090A Field Master Pro is a synthesizer-based handheld spectrum analyzer that provides quick and accurate measurement results. The instrument is designed for monitoring, measuring, and analyzing signal environments. Measurements can easily be made by using the main instrument functions: frequency, span, amplitude, and bandwidth.
  • Page 10 Vision High-Speed Port Scanner Enabled MS2090A-0421* Pulse Analyzer MS2090A-0431* Coverage Mapping (requires GPS option MS2090A-0031) MS2090A-0444* EMF Measurement (requires a compatible Anritsu isotropic antenna) MS2090A-0445* EMF Meter Enabled (requires 2000-1985-R isotropic EMF probe) MS20x0A-0509* AM/FM Modulation Measurement MS2090A-0883* LTE FDD/TDD Measurement (requires GPS option MS2090A-0031)
  • Page 11: Instrument Care And Preventive Maintenance

    General Information 1-3 Instrument Care and Preventive Maintenance Instrument Care and Preventive Maintenance Instrument care and preventive maintenance consist of proper operation in a suitable environment, occasional cleaning of the instrument, and inspecting and cleaning the RF connectors and all accessories before use. Clean the instrument with a soft, lint-free cloth dampened with water or water and a mild cleaning solution.
  • Page 12: Esd Caution

    Anritsu recommends annual calibration and performance verification by local Anritsu service centers. Accredited calibration to ISO17025 and ANSI/NCSL Z540-1 are available and can include a calibration certificate, test report, and uncertainty data. Contact Anritsu sales and service centers for more information. Contacting Anritsu for Sales and Service To contact Anritsu, visit the following URL and select the services in your region: http://www.anritsu.com/contact-us...
  • Page 13: Chapter 2-Instrument Overview

    Chapter 2 — Instrument Overview Introduction This chapter provides an overview of the Anritsu MS2090A Field Master Pro. It describes the instrument hardware features, touch screen display, general system settings, GUI overview and instrument configurations, and the connector panels. For detailed information on the instrument’s user interface and functions, refer to the relevant measurement guide.
  • Page 14: Instrument Front Panel

    2-2 Instrument Front Panel Instrument Overview Instrument Front Panel The Field Master Pro uses a touch screen for data input. The menus can vary depending on the current measurement configuration, installed options, and selected instrument function. 1. LCD Touch Screen Display “Top Connector Panel”...
  • Page 15: Connector Panels

    Instrument Overview 2-3 Connector Panels Connector Panels The MS2090A Field Master Pro uses two connector panels to provide for all physical IO. These panels use a variety of connector types intended for their purpose. RF Connectors The main RF input connector is Type N, Type K, or Type V, depending on the frequency option that is installed.
  • Page 16 2-3 Connector Panels Instrument Overview 2. IF Out: This 50  SMA connector provides internal IF output. The nominal frequency range is 325 MHz with FFT capture bandwidth ≤ 32 MHz and 300 MHz with FFT capture bandwidth = 100 MHz. The nominal RF output level is –4 dBm with a 10 MHz, –20 dBm input, and with 0 dB input attenuation and preamp off.
  • Page 17: Side Connector Panel

    Instrument Overview 2-3 Connector Panels Side Connector Panel Figure 2-3 shows the MS2090A side connector panel. Data Figure 2-3. Side Connector Panel The MS2090A is compatible with external USB memory devices that have an integrated keypad and Note are FIPS compliant using AES 256-bit encryption. 1.
  • Page 18 2-3 Connector Panels Instrument Overview 5. MicroSD: The Micro Secure Digital slot is a small expansion slot located on the side panel. The slot accepts industry standard MicroSD storage cards and can be used for storing measurements, setup data, and screen images similar to USB storage devices.
  • Page 19: Tilt Bail Stand

    Instrument Overview 2-4 Tilt Bail Stand Tilt Bail Stand The attached tilt bail can be used for desktop operation. The tilt bail provides a backward tilt for improved stability and air flow. To deploy the tilt bail, pull the bottom of the tilt bail away from the back of the instrument.
  • Page 20: Battery Information

    If the adapter plug becomes hot to the touch during operation, discontinue use immediately. Anritsu Company recommends removing the battery for long-term storage of the instrument. The batteries will charge at a faster rate when the instrument is turned off.
  • Page 21: Replacing The Battery

    Instrument Overview 2-5 Battery Information Replacing the Battery The battery can be replaced without the use of tools. The battery compartment door is located on the lower right side of the instrument (when you are facing the measurement display). To remove the battery: Push in and slide the battery door tab to disengage it.
  • Page 22: Turning On The Ms2090A Field Master Pro

    2-6 Turning On the MS2090A Field Master Pro Instrument Overview Turning On the MS2090A Field Master Pro The Anritsu MS2090A Field Master Pro is capable of approximately two hours of continuous operation from a fully charged, field-replaceable battery (refer to “Battery Information” on page 2-8).
  • Page 23: Gui Overview

    Instrument Overview 2-7 GUI Overview GUI Overview The MS2090A Field Master Pro software controls all instrument functions. The software runs locally on the instrument and primary operation is through the touch screen display. The figure below identifies the main display areas, which are each described in more detail later in this chapter. 1.
  • Page 24: Operating The Touch Screen

    2-7 GUI Overview Instrument Overview Operating the Touch Screen Field Master Pro uses common touch gestures to achieve a variety of operations. These include interactions on hot areas by pressing, double pressing (or tapping), dragging, and pinching. Single Press: Most controls require a single press or tap.
  • Page 25: Title Bar

    Instrument Overview 2-7 GUI Overview Title bar The tool bar is located at the top of the interface screen. It displays icons that provide access to information and user actions as described below. All of the icons are active and will open the appropriate menu or item when selected.
  • Page 26 2-7 GUI Overview Instrument Overview Common GUI Controls In addition to the touch gestures described in the previous section, the following table lists the most common controls displayed in the toolbar of Field Master Pro. The 3-line icon provides quick access to system information, settings, file management, and built-in diagnostic tools.
  • Page 27: Data Entry

    Instrument Overview 2-7 GUI Overview Select the padlock icon next to center frequency on the graticule to change the frequency by dragging, when using a touch-screen enabled PC. Select the padlock icon next to SPAN to change the span by pinching, when using a touch-screen enabled PC.
  • Page 28 2-7 GUI Overview Instrument Overview A physical keyboard connected via USB can also be used to enter values in a similar manner and the keyboard ENTER key used to accept the new value. Selection Lists Some parameters and instrument functions are selectable from a list. These list boxes display the available selections and value limits as applicable.
  • Page 29: Scroll Indication

    Instrument Overview 2-7 GUI Overview Scroll Indication In some cases, menus, status panels, or other lists contain more information than can be displayed in the available area. In these cases, the top or bottom of the panel will have a fade-to-white appearance as shown below.
  • Page 30 2-7 GUI Overview Instrument Overview Text Entry and EZ Keyboard When an instrument function requires you to enter text, such as entering an Ethernet hostname or IP address, a touch screen alphabetic keyboard is displayed. See Figure 2-13. Select the “?123” key to switch to the digits and symbols keyboard.
  • Page 31 Instrument Overview 2-7 GUI Overview To change a key’s value: Select the pencil (edit) key. The EZ keys will become highlighted. Figure 2-15. Touch Screen EZ Keyboard Select the key to be edited. This will display the standard keyboard for entering a new EZ key value. Figure 2-16.
  • Page 32: Selecting The Analyzer

    2-8 Selecting the Analyzer Instrument Overview Selecting the Analyzer The instrument analyzers are selected from the 9-dot icon. To select an analyzer, select the 9-dot icon in the Title bar to display the available analyzers, illustrated in Figure 2-17. Simply select the desired icon to load the new analyzer.
  • Page 33: System Menu

    From here you can also toggle the WEB UPDATES button to automatically install the latest software update when it is available on the Anritsu website. This functionality facilitates firmware updates within the last 24 hours, if the instrument is connected to the Internet.
  • Page 34: Notifications Icon

    2-9 System Menu Instrument Overview Notifications Icon Select Notifications icon to view all dynamic notifications of the instrument. These notifications consists of a mixture of informational, warning and errors messages. Refer to Appendix A, “Instrument Notifications and Troubleshooting” for detailed information. Figure 2-20.
  • Page 35: 2-10 Settings Menu

    Instrument Overview 2-10 Settings Menu 2-10 Settings Menu The system settings menu provides access to all instrument system-level settings such as network, GNSS/GPS, date and time, and display. DISPLAY: Opens the Display Settings. SOUND: Opens the Sound Settings. ETHERNET: Opens the Network Settings.
  • Page 36: Display Settings

    2-10 Settings Menu Instrument Overview Display Settings The DISPLAY settings allow you to adjust the display brightness, choose the display color scheme and to view or hide shortcuts to the saved application settings. Brightness The BRIGHTNESS option consists of a slider to adjust the screen brightness, in addition to selecting the amount of idle time before the screen automatically dims to conserve battery life.
  • Page 37 Instrument Overview 2-10 Settings Menu Theme Under THEME option select COLOR SCHEME drop-down menu and select either Default, Light, Black on White or Night Vision to set the display color scheme of your choice. If the THEME is set to Light it may be more suitable for viewing in bright ambient conditions.
  • Page 38 2-10 Settings Menu Instrument Overview Figure 2-24. Black on White Theme Display Figure 2-25. Night Vision Theme Display 2-26 PN: 10580-00444 Rev. U MS2090A UG...
  • Page 39 Instrument Overview 2-10 Settings Menu Shortcuts Under the SHORTCUTS option, toggle the HIDE SHORTCUTS button to view or hide the shortcuts. Select + SHORTCUT button located at the bottom left of the display screen to create or save specific settings of an application.
  • Page 40 2-10 Settings Menu Instrument Overview Figure 2-28. Add Shortcut Dialog Enter a file name to save the current setup, and select SAVE. Figure 2-29. File Save As Select the newly created shortcut to recall the saved setup file. To delete all the shortcuts, select SYSTEM MENU (3-line icon), go to SETTINGS menu, select DISPLAY settings and select CLEAR ALL button at the bottom right of the instrument’s display.
  • Page 41: Sound Settings

    Instrument Overview 2-10 Settings Menu Sound Settings The SOUND settings allow you to adjust the SYSTEM VOLUME or MUTE ALL of the instrument sounds. The system volume may affect other volume settings found in various setup menus. Figure 2-31. Sound Settings Network Settings The MS2090A uses Ethernet or WLAN (Wi-Fi) to communicate remotely with a controller.
  • Page 42 2-10 Settings Menu Instrument Overview • The remote program may connect to the instrument IP address or to its HOSTNAME (Ethernet only). If using DHCP instead of a static IP, using the HOSTNAME may be more reliable for finding an instrument on a network.
  • Page 43 Instrument Overview 2-10 Settings Menu The instrument IP address and its HOSTNAME are set via the System menu (upper left corner) and accessing the ETHERNET or WIFI settings menu. Wi-Fi does not support connections using HOSTNAME; use IP addressing to establish a wireless Note network connection.
  • Page 44: Ethernet Settings

    2-10 Settings Menu Instrument Overview Ethernet Settings Refer to “Network Settings” on page 2-29 for general network setup and information. Access the System menu (3-line icon in the upper left corner). Select SETTINGS to access the instrument settings menu, then select ETHERNET to view the current network settings (IP address, HOSTNAME, etc.).
  • Page 45: Wi-Fi Settings

    Instrument Overview 2-10 Settings Menu Wi-Fi Settings Refer to “Network Settings” on page 2-29 for general network setup and information. Access the System menu (3-line icon in the upper left corner). Select SETTINGS to access the instrument settings menu, then select WIFI to display the current network settings (IP address, HOSTNAME, etc.).
  • Page 46: Gnss (Gps) Settings (Option 31)

    2-10 Settings Menu Instrument Overview GNSS (GPS) Settings (Option 31) The MS2090A Field Master Pro is available with a built-in global positioning receiver feature (Option 31) that can provide latitude, longitude, altitude, and UTC timing information. This option also enhances frequency reference oscillator accuracy.
  • Page 47 Instrument Overview 2-10 Settings Menu When the GNSS (GPS) receiver has established a “good fix”, the GNSS (GPS) icon is displayed with a center dot and the following information is kept updated: • Fix status • Tracked satellites • Latitude •...
  • Page 48: Screenshot Setup

    2-10 Settings Menu Instrument Overview Screenshot Setup Access the System menu (3-line icon in the upper left corner). Select SETTINGS to access the instrument settings menu, then select SCREENSHOT to open the screenshot setup menu. Figure 2-35. Screenshot Setup Screenshot setup lets you configure the following: •...
  • Page 49: Language Settings

    Instrument Overview 2-10 Settings Menu Language Settings Follow the steps below to access language settings in order to change a selected display language: Access the System menu (3-line icon in the upper left corner). Select SETTINGS to access the instrument settings menu, then select LANGUAGE to open the language menu.
  • Page 50: Date Time Settings

    2-10 Settings Menu Instrument Overview Date Time Settings Access the System menu (3-line icon in the upper left corner). Select SETTINGS to access the instrument settings menu, then select DATE TIME to display the current date and time settings. Figure 2-37. Date Time Settings The Date Time setup lets you set the current date and time and the time zone.
  • Page 51: Port Setup

    Instrument Overview 2-10 Settings Menu Port Setup The Port Setup menu allows you to configure the external ports. Bias Voltage Bias Voltage setup provides controls to set the voltage level and to monitor the precise voltage and current, and trip state. Access the System menu (3-line icon in the upper left corner).
  • Page 52 2-10 Settings Menu Instrument Overview Reference and Trigger The reference and trigger setup provides controls to set the functionality of the configurable top panel reference and trigger ports. Access the System menu (3-line icon in the upper left corner). Select SETTINGS to access the instrument settings menu, then select PORT SETUP > REF / TRIG to display the top panel reference and trigger port settings.
  • Page 53: Options Settings

    WEB to install options available on the web. Ensure that the unit is connected to the Internet. Select ENABLE OPTIONS USING FILE to install options file saved in USB memory device. Select SAVE CONFIG to export the config file to Anritsu. Figure 2-40. Options Settings...
  • Page 54 2-10 Settings Menu Instrument Overview To activate the installed time-limited options select the ACTIVATE button to begin the 90-day trial. Figure 2-41. Options Settings 2-42 PN: 10580-00444 Rev. U MS2090A UG...
  • Page 55: Maps Settings (Option 431)

    Instrument Overview 2-10 Settings Menu Maps Settings (Option 431) Map settings are available when Option 431, Coverage Mapping is installed. Maps settings are used to delete map tiles from instrument memory. Refer to Coverage Mapping section in spectrum analyzer measurement guide (10580-00447).
  • Page 56: Advanced Settings

    2-10 Settings Menu Instrument Overview Advanced Settings Advanced settings provides an RF SAFE MODE toggle to protect the internal detection circuitry on boot-up. Access the System menu (3-line icon in the upper left corner). Select SETTINGS to access the instrument settings menu. Select ADVANCED to enable/disable the RF SAFE MODE and SCPI ERRORS controls.
  • Page 57: Reset Settings

    Instrument Overview 2-10 Settings Menu Reset Settings Access the System menu (3-line icon in the upper left corner). Select SETTINGS to access the instrument settings menu. Select RESET to open the Reset menu. Figure 2-44. Reset Settings The following reset options are available: SETTINGS RESET •...
  • Page 58: File Management

    2-11 File Management Instrument Overview 2-11 File Management Access the System menu (3-line icon in the upper left corner). Select FILES to access the instrument file management menu. Figure 2-45. File Management Menu 2-46 PN: 10580-00444 Rev. U MS2090A UG...
  • Page 59: File Locations

    Instrument Overview 2-11 File Management File Locations Displays the available memory locations. Select the location that you want to access and the available files will be displayed on the right side. You can select the column headers to change the sort order. File Management Operations All file operations are selected via the following icons: 3-Dots...
  • Page 60: Previewing Screenshots

    2-11 File Management Instrument Overview Previewing Screenshots While in File Manager, you can preview screenshots on the instrument display simply by selecting the screenshot file name. While the preview is displayed, you can use the controls to: • Cycle forward and backward through each screenshot in the directory. •...
  • Page 61: 2-12 Diagnostics Menu

    Instrument Overview 2-12 Diagnostics Menu 2-12 Diagnostics Menu Select System menu. Select DIAGNOSTICS menu to access the diagnostics options of the instrument. See Figure 2-48. BATTERY: Displays the list of battery information. Refer to Battery section. EVENT LOG: Saves log files in a USB memory device. Refer to Event Log section.
  • Page 62: Event Log

    2-12 Diagnostics Menu Instrument Overview Event Log The event log allows you to save the instrument notifications that are not displayed in the title bar. Select the SAVE button to copy the event log file into a USB memory device. This event log file is a binary file which is used by the customer service for troubleshooting purposes.
  • Page 63: Service Mode

    Instrument Overview 2-12 Diagnostics Menu Service Mode The service mode is for Anritsu customer service use only. Figure 2-52. Service Mode MS2090A UG PN: 10580-00444 Rev. U 2-51...
  • Page 64: 2-13 Tools Menu

    2-13 Tools Menu Instrument Overview 2-13 Tools Menu Access the System menu (3-line icon in the upper left corner). Select TOOLS to access the instrument tools menu, then select the tool menu that you wish to open. WEB: Opens “WEB” tool.
  • Page 65: Iq Streaming

    Instrument Overview 2-13 Tools Menu IQ Streaming The IQ STREAMING menu allows you to concatenate an existing IQ metadata file (.dgzm) with a session of IQ data files (.dgz) that were streamed to a USB device after the stream operation is complete. The data files must all come from the same stream session.
  • Page 66: Pdf Reports

    2-13 Tools Menu Instrument Overview PDF REPORTS The PDF REPORTS tool allows the user to generate a PDF report or HTML report. Select OPEN REPORT GENERATOR to open REPORT GENERATOR – CONFIG window. Figure 2-56. PDF Reports Tool In REPORT GENERATOR – CONFIG window you can load a template to populate the given fields. •...
  • Page 67 Instrument Overview 2-13 Tools Menu Figure 2-57. Report Generator - Config Window Select SETUP button in the bottom to add setup or trace files in REPORT GENERATOR - SETUP window. • Select the blue + ADD FILE (S) button in the upper right to add image files (.png. jpeg, jpg, .fmcaausb) to the PDF report.
  • Page 68 2-13 Tools Menu Instrument Overview Figure 2-58. Report Generator - Setup Window Select PREVIEW button to see the preview of the generated report in REPORT GENERATOR - PREVIEW window. 2-56 PN: 10580-00444 Rev. U MS2090A UG...
  • Page 69 Instrument Overview 2-13 Tools Menu Select SAVE TO PDF button on the top to save the report internally in the REPORTS folder. If you want to preview an internally saved HTML file, select the 3-dot icon in the top right of the preview window.
  • Page 70 2-13 Tools Menu Instrument Overview 2-58 PN: 10580-00444 Rev. U MS2090A UG...
  • Page 71: Appendix A-Instrument Notifications And Troubleshooting

    Troubleshooting Introduction This appendix provides a list of instrument messages and their meaning and procedures for a full system recovery. If any error condition persists, contact your local Anritsu Service Center (http://www.anritsu.com/contact-us). Field Master Pro displays the following message types: •...
  • Page 72: Self-Test Messages

    A-2 Self-Test Messages Instrument Notifications and Troubleshooting Self-Test Messages Field Master Pro has built-in diagnostics that allow you to run a built-in self test. The test results can be viewed on the screen and saved to a log file. To run self test, use the 3-line menu to access DIAGNOSTICS, then select SELF TEST and RUN SELF TEST.
  • Page 73 Instrument Notifications and Troubleshooting A-2 Self-Test Messages Motherboard Temperature : 46.5 C Fan speed, percent of full speed : 7 % Fan speed : 2638 RPM Temperature at Reference : 45.3 C SPA : OVERALL STATUS : PASS +12VG : 12.00 V, 1.20 V ,11.90 V (PASS) +5V_PREAMP : 5.00 V, 0.50 V ,4.99 V (PASS)
  • Page 74 ADC Thermal Sensor : 35.00 C, 90.00 C ,39.93 C (PASS) If any self test fails, try resetting the instrument with a RESET ALL and reboot, and run the test again. If the condition persists, contact your local Anritsu Service Center (http://www.anritsu.com/contact-us). PN: 10580-00444 Rev. U MS2090A UG...
  • Page 75: File Management Notifications

    File Management Notifications File management functions (accessed by via System menu (3-line icon) > FILES) are used to organize, copy, and rename files. The following notifications are displayed in a persistent dialog when using the file manager features. Cannot access File or Directory User does not have permission to access the file or directory.
  • Page 76: Informational Notifications

    Informational Notifications Audio must be turned on to record. In Toolbar, Not Logged, Transient Battery Over Charge Temperature. In Toolbar, Not Logged, Persistent Cannot start measurement without a connected Antenna. In Toolbar, Not Logged, Persistent Calibration abort is not allowed. In Toolbar, Not Logged, Transient Calibration temperature range exceeded.
  • Page 77 External reference disconnected – switching reference source to GPS In Toolbar, Logged, Transient External reference disconnected – switching reference source to internal In Toolbar, Logged, Transient External reference disconnected – switching reference source to Internal High In Toolbar, Logged, Transient File Does Not Exist In Toolbar, Not Logged, Transient File Recall Failed...
  • Page 78 A-4 Informational Notifications Instrument Notifications and Troubleshooting GPS reference disconnected – switching reference source to Internal In Toolbar, Logged, Transient GPS reference no longer available – switching reference source to Internal High In Toolbar, Logged, Transient IF Gain Cal completed In Toolbar, Logged, Transient Internal High reference no longer valid –...
  • Page 79 Instrument Notifications and Troubleshooting A-4 Informational Notifications Measurement changed. Trace playback stopped In Toolbar, Logged, Transient Measurement changed. Trace recording in csv stopped In Toolbar, Logged, Transient Memory Full Popup, Not Logged, Transient Memory Full In Toolbar, Not Logged, Transient Model and/or serial number of the device does not match calibration stored.
  • Page 80 A-4 Informational Notifications Instrument Notifications and Troubleshooting Presetting Instrument In Toolbar, Not Logged, Transient Rebooting Instrument In Toolbar, Not Logged, Transient Rebooting instrument into Recovery mode In Toolbar, Not Logged, Transient Recall initiated When a file is selected for recall Reset Completed In Toolbar, Logged, Transient Resetting Instrument...
  • Page 81 Instrument Notifications and Troubleshooting A-4 Informational Notifications <Name of Setting> range adjusted to accommodate user change In Toolbar, Not logged, Transient. The setting can be distance, time and frequency. Self test Passed In Toolbar, Logged, Transient Self test Failed In Toolbar, Logged, Transient Software update files found on USB In Toolbar, Not Logged, Transient Sensor <model>...
  • Page 82 A-4 Informational Notifications Instrument Notifications and Troubleshooting Zeroing. Please wait In Toolbar, Logged, Persistent Zeroing <sensor> In Toolbar, Not Logged, Persistent Zeroing Sensor ... In Toolbar, Logged, Persistent Zeroing Sensor completed In Toolbar, Logged, Transient No sensor connected In Toolbar, Not Logged, Persistent USB sensor <model>...
  • Page 83: Warning Notifications

    Instrument Notifications and Troubleshooting A-5 Warning Notifications Warning Notifications ADC Overrange In Toolbar, Logged, Persistent Antenna information invalid In Toolbar, Not Logged, Persistent Battery Fault In Toolbar, Logged, Transient Battery level is critically low. Automatic shutdown is imminent Popup, Logged, Persistent Battery level low Popup, Logged, Persistent Battery Low warning...
  • Page 84 A-5 Warning Notifications Instrument Notifications and Troubleshooting Failed to save screenshot. Device may be full. When saving a screenshot fails. IF Gain Calibration is off In Toolbar, Logged, Persistent IF Shape Calibration is off In Toolbar, Logged, Persistent Invalid frequency setup for this antenna In Toolbar, Logged, Transient Limit recall initiated When a limit setup file is selected for recall.
  • Page 85 Instrument Notifications and Troubleshooting A-5 Warning Notifications Setting change detected; unloading playback file In Toolbar, Logged, Transient Save failed. File size exceeds save limits. Adjust setup and try again In Toolbar, Logged, Transient Shutdown Battery Level Reached Pop-up Window, Logged, Persistent: Requires User Intervention Sent after a Battery Low warning.
  • Page 86 A-5 Warning Notifications Instrument Notifications and Troubleshooting Quasi-peak detection selected. Sweep will be slower In Toolbar, Logged, Transient Zero Span is not available In Toolbar, Logged, Transient <Name of setting> <Value of setting> exceeds the min/max range <minimum value - maximum value> In Toolbar, Logged, Transient <setting>...
  • Page 87: Error Notifications

    Instrument Notifications and Troubleshooting A-6 Error Notifications Error Notifications Antenna disconnected. Terminating measurement In Toolbar, Logged, Transient ALC Unlevel Error (X) @ Y Hz In Toolbar, Logged, Transient. X represents the ALC that failed and Y represents the specific frequency value at which the failure occurred.
  • Page 88 A-6 Error Notifications Instrument Notifications and Troubleshooting IQ capture file failed to save In Toolbar, Logged, Transient IQ Capture Aborted In Toolbar, Logged, Transient Model Mismatch. Shortcut Creation failed due to model mismatch. Model from file. In Toolbar, Logged, Transient Not enough available space to save file: filepath/filename When the entire file could not be written Out of Range...
  • Page 89 Instrument Notifications and Troubleshooting A-6 Error Notifications The selected custom mask is invalid In Toolbar, Logged, Transient Trace playback not available In Toolbar, Logged, Transient Trace recording not available In Toolbar, Logged, Transient Trace has not been copied to memory In Toolbar, Logged, Transient Unexpected software exception.
  • Page 90: System Recovery

    Some features of RECOVERY MODE are not available in instruments that have been upgraded from firmware versions prior to 2021.11.1. This affects using SLOT B as the boot position or the target for Note a firmware update. These instruments must be returned to an Anritsu service center to update the recovery image. Visit www.anritsu.com/contact-us to find your nearest Anritsu Service and Support Center.
  • Page 91 Instrument Notifications and Troubleshooting A-7 System Recovery The Recovery Mode menu (Figure A-2) shows instrument information and provides a variety of recovery options in ascending invasive order: Select REBOOT to reboot the instrument without any further action. Save all internal user files and logs by inserting a USB device and press the EXPORT button. Refer to “Reset”...
  • Page 92: Reset

    A-7 System Recovery Instrument Notifications and Troubleshooting Reset The recovery RESET dialog is shown in Figure A-3. Here you can reset the MS2090A from the choices shown. In rare cases, system, internal, or other user files may interfere with instrument operation, so choosing to delete this data in that order may resolve a problem.
  • Page 93: Install Sw To Slot X

    Instrument Notifications and Troubleshooting A-7 System Recovery Install SW to SLOT X The recovery mode shows two slots (A and B). These slots contain different system software packages that are installed. Each slot contains the following information: • Slot name (A,B): Boot version package (A or B). •...
  • Page 94: Udp Discovery Tool Firewall Settings

    A-8 UDP Discovery Tool Firewall Settings Instrument Notifications and Troubleshooting UDP Discovery Tool Firewall Settings Follow the instructions below if the Discovery tool is not responding: Go to Control Panel and open Windows Defender Firewall settings. Click Allow an app or feature through Windows Defender Firewall. Figure A-5.
  • Page 95 Instrument Notifications and Troubleshooting A-8 UDP Discovery Tool Firewall Settings Find fieldmaster_base in the list of apps, and make sure that the fieldmaster_base box and Domain box are checked. Private and Public boxes may also need to be checked depending on where the instruments are on the network.
  • Page 96: System Override

    A-9 System Override Instrument Notifications and Troubleshooting System Override In the event that the instrument does not respond to SCPI commands (due to circumstances like a long sweep used in conjunction with *OPC?) and needs to be reset remotely, the system override feature can be used. System override can be accessed through TCP port 8001 of the instrument.
  • Page 97: Appendix B-Upgrading Software Options

    The instrument can be upgraded to software options in either of the two ways, Enable Options Using File and Install Options From Web: After purchasing a software option from Anritsu, place the license file on the USB memory device in the top (root) level, not in any directory.
  • Page 98: Software Update

    Upgrading Software Options Software Update To update your Anritsu instrument software, use a high quality USB memory device with at least 1 GB of free space and FAT32 file system format. Insert the USB memory device into your PC or laptop.
  • Page 99 Upgrading Software Options B-4 Software Update Verify the Software Update Once the software update is complete, power off the MS2090A. Power on the MS2090A. Select the System menu icon (3-line icon) from the MS2090A to display the System Menu. Select the System Information Menu menu (see Figure...
  • Page 100 B-4 Software Update Upgrading Software Options PN: 10580-00444 Rev. U MS2090A UG...
  • Page 101: Appendix C-Pc Software

    Appendix C — PC Software Introduction The MS2090A remote user interface software provides remote access to a network connected instrument and displays the same MS2090A user interface, instrument controls, and live measurement data on the computer screen. The software also allows you to load measurement and setup data from saved files and then perform measurement analysis on the recalled traces, even when instrument hardware is not available or is not connected to the computer by using the localhost (refer to Section C-4 “Connecting to the MS2090A or...
  • Page 102: Installation

    C-3 Installation PC Software Installation Download the MS2090A software from the product page: http://www.anritsu.com/en-US/test-measurement/products/ms2090a Launch the MS2090A executable file and follow the on-screen instructions after selecting YES to the User Account Control dialog. Figure C-1. MS2090A PC Software Installation PN: 10580-00444 Rev. U...
  • Page 103: Connecting To The Ms2090A Or Localhost

    PC Software C-4 Connecting to the MS2090A or Localhost Connecting to the MS2090A or Localhost Connecting to a networked instrument or to the localhost is established via the system menu. If connecting to an instrument, the instrument must first be connected to the network via Ethernet or Wi-Fi and the IP address of the instrument must be known.
  • Page 104: System Information

    C-5 System Menu PC Software System Information Access the System menu (3-line icon in the upper left corner). Click SETTINGS to access the settings menu, then select SYSTEM INFORMATION to view the system details. Figure C-3. System Information Menu PN: 10580-00444 Rev. U MS2090A UG...
  • Page 105: Files (File Management

    PC Software C-5 System Menu Files (File Management) The PC file system is used for all file operations when connected to either an instrument or to the localhost. The FILE menu will display the Windows Save As or Recall File dialogs; the FILES system menu will launch Windows Explorer.
  • Page 106: Remote Files

    C-5 System Menu PC Software Remote Files Access the System menu (3-line icon in the upper left corner). Click SETTINGS to access the settings menu, then select REMOTE FILES to open the default browser window and download the internally saved files for e.g. screenshots. See Figure C-5.
  • Page 107: Tools Menu

    PC Software C-5 System Menu Tools Menu Access the System menu (3-line icon in the upper left corner). Click TOOLS to access the tools menu, then select the tool menu that you wish to open. IQ STREAMING: Opens IQ Streaming Tool. Refer to IQ Streaming Tools section of Field Master Series Spectrum Analyzer Measurement Guide (10580-00447).
  • Page 108: Working With The Localhost

    C-6 Working with the Localhost PC Software Working with the Localhost The localhost setting provides a simulated connection to the instrument hardware, but does not provide simulated data. When an instrument is not connected and the MS2090A software instead connects with the localhost, most of the menus are available and operate the same as they do on the instrument, but there are some differences as discussed in the following sections.
  • Page 109 PC Software C-6 Working with the Localhost FREQ SPAN Changes here suppress recalled data. AMPLITUDE Changes to REF LEVEL, SCALE/DIV, Y AXIS UNIT will change the graticule. PRE AMP and ATTEN LEVEL have no affect. All other changes will suppress recalled data. BANDWIDTH Changes here suppress recalled data.
  • Page 110 C-6 Working with the Localhost PC Software C-10 PN: 10580-00444 Rev. U MS2090A UG...
  • Page 112 Anritsu Company 490 Jarvis Drive Anritsu utilizes recycled paper and environmentally conscious inks and toner. Morgan Hill, CA 95037-2809 http://www.anritsu.com...

Table of Contents