pH 1100 L
HOLD function
Overview
Settings
ba77068d05
05/2021
Measured
parameter
Temperature
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
1
Freeze the measured value with
The [HOLD] status indicator is displayed.
The HOLD function is active.
2
Release the frozen measured value again with
<MODE>
.
The [HOLD] status display disappears. The display switches
back to the previous indication.
5.4.4 Settings for pH and ORP measurements
The following settings are possible for pH and ORP measurements:
Resolution
Calibration interval
Buffers for calibration
Unit of the temperature
Automatic stability control
Unit for slope
Calibration record (display)
The settings are made in the menu for calibration and measurement
settings of the pH/ORP measurement. To open the settings, display the
required parameter in the measured value display and press the
<OK>
[menu] or
key. After completing the settings, switch to the mea-
sured value display with
Menu item
Calibration / Calibra-
tion record
Time interval
Stability during the time
interval
Δ : better than 0.5 °C
15 seconds
<HOLD>
<MODE>
.
Possible
Explanation
setting
-
Displays the calibration
record of the last calibra-
tion.
Operation
.
<HOLD>
or
<F1>
31
/