pH 1100 L
ba77068d05
05/2021
1
In the measured value display, select the measured parameter
<MODE>
pH or mV with
2
Start the calibration with
The calibration display appears.
pH
0.0
24.8 °C
AnyCal
15.03.2014
08:00
3
Thoroughly rinse the electrode with deionized water.
4
Immerse the electrode in buffer solution 1.
5
When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
<
><
>
with
6
Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
pH
0.0
24.8 °C
AnyCal
AR
15.03.2014
08:00
7
Wait for the end of the measurement with stability control or
accept the calibration value with
.
<CAL>
.
1
Buffer
mV
.
<OK>
.
7.000
Buffer
mV
<OK>
Operation
.
41