Firmware update
Available firmware updates are provided on the Internet.
With the "Firmware Update " program and a PC you can update the
firmware of the pH 1100 L to the newest version.
For the update you have to connect the meter to a PC.
For the update via the USB interface, the following is required:
 a free USB interface (virtual COM port) on the PC
 the driver for the USB interface (on the enclosed CD-ROM)
 the USB cable (included in the scope of delivery of the pH 1100 L).
1
Install the downloaded firmware update on a PC.
An update folder is created in the Windows start menu.
If an update folder already exists for the meter (or meter type),
the new data is displayed there.
2
In the windows start menu, open the update folder and start the
firmware update program.
3
Using the USB interface cable, connect the pH 1100 L to a
USB interface (virtual COM port) of the PC.
4
Switch on the pH 1100 L.
5
In the firmware update program, start the update process with
OK.
6
Follow the instructions of the firmware update program.
During the programming process, a corresponding message
and a progress bar (in %) are displayed.
The programming process takes approx. three minutes. A ter-
minatory message is displayed after a successful programming
process. The firmware update is completed.
7
Disconnect the pH 1100 L from the PC.
The pH 1100 L is ready for operation again.
After switching the meter off and on you can check whether the meter
has taken over the new software version (see page 63).
pH 1100 L