Operation
Stability control
(AutoRead )
Criteria for a stable
measured value
30
2
Release the frozen measured value again with
<MODE>
.
The [HOLD] status display disappears. The display switches
back to the previous indication.
5.4.3 Measuring the ORP
Note
ORP electrodes are not calibrated. However, you can check ORP elec-
trodes using a test solution.
1
Perform the preparatory activities according to section 5.4.1.
2
Submerse the ORP electrode in the sample.
U
157.0
24.8 °C
15.03.2014
Menu
08:00
3
Select the mV display with
The stability control function (AutoRead) continually checks the stability
of the measurement signal. The stability has a considerable impact on
the reproducibility of measured values.
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic Stability control is switched off.
The Stability control function checks whether the measured values are
stable within the monitored time interval.
Measured
parameter
ORP
mV
USB output
<MODE>
Time interval
Stability during the time
interval
Δ : better than 0.3 mV
15 seconds
pH 1100 L
<HOLD>
.
ba77068d05
or
05/2021