Operation
Sample record
Calibration evaluation
36
30.03.2014 15:55
pH 1100 L
Ser. no. 08502113
CALIBRATION pH
AutoCal TEC
Buffer 1
Buffer 2
Buffer 3
Voltage 1
Voltage 2
Voltage 3
Temperatur 1
Temperatur 2
Temperatur 3
Slope
Asymmetry
Sensor
etc...
After calibrating, the meter automatically evaluates the calibration.The
zero point and slope are evaluated separately. The worse evaluation of
both is taken into account. The evaluation appears on the display and
in the calibration record.
Display
Clean the electrode according to the
electrode operating manual
Error
Eliminate the error according to chapter
7 W
... (page 61)
HAT TO DO IF
4.00
7.00
10.00
184.0 mV
3.0 mV
-177.0 mV
24.0 øC
24.0 øC
24.0 øC
-60.2 mV/pH
4.0 mV
+++
Calibration
Zero point
record
[mV]
+++
-15 ... +15
++
-20 ... +20
+
-25 ... +25
-
-30 ... +30
< -30 or
Error
> 30
pH 1100 L
Slope
[mV/pH]
-60,5 ... -58
-58 ... -57
-61 ... -60.5
or
-57 ... -56
-62 ... -61
or
-56 ... -50
... -62 or
... -50
ba77068d05
05/2021