TSI Instruments VelociCalc 9600 Series Operation And Service Manual page 35

Air velocity meter
Table of Contents

Advertisement

Calibration Offset Adjustments Applicable Models
Temperature
Barometric Pressure
960, 962,
9600, 9630, 9650,
964, 966,
9650-NB
980, 982,
800220
You can apply offset adjustments
to temperature, relative humidity,
and barometric pressure. The
measurement displayed and
logged will be the raw
measurement added to the value
you specify here. Offsets will be
shown on the dashboard if the
value is other than 1. Enter 0 to
return to default.
Correction Factors Applicable Models
Pitot Velocity
9630, 9650, 9650-
9630, 9650, 9650-NB
NB
You can apply correction factors to
pitot velocity, pitot flow and thermal
anemometer and rotating vane air
flows. The measurement displayed
and logged will be the raw
measurement multiplied by the
value you specify here. Correction
factors will be shown on the
dashboard if the value is other than
1. Enter 1.000 to return to default.
IAQ Probe Calibration Applicable Models
Carbon Dioxide (CO
)
2
980, 982
Operation and Service Manual
Relative Humidity
964, 966, 980, 982, 995,
800220
Pitot Flow
960, 962, 964, 966,
995
Carbon Monoxide (CO)
982
Probe Flow
25