Summary of Contents for TSI Instruments VelociCalc 9600 Series
Page 1
VelociCalc ® VelociCalc ® Air Velocity Meter Model 9600 Series Operation and Service Manual P/N 6016122, Revision C August 2022 www.tsi.com...
Page 2
Start Seeing the Benefits of Registering Today! ® ® Thank you for your TSI instrument purchase. Occasionally, TSI releases information on software updates, product enhancements and ® new products. By registering your instrument, TSI will be able to send this important information to you. http://register.tsi.com As part of the registration process, you will be asked for your comments on TSI products and services.
Warranty Copyright TSI Incorporated / 2021-2022 / All rights reserved. Address TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA Fax No. 651-490-3824 E-mail Address answers@tsi.com Limitation of Warranty (For country-specific terms and conditions outside of the USA, please visit and Liability www.tsi.com.) (effective February 2015)
Page 4
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE GOODS.
Contents WARRANTY ..................III SAFETY .................... VII Description of Caution/Warning Symbols ........vii Caution ..................vii Warning ..................vii Caution and Warning Symbols ............viii Labels ..................... viii RoHS ....................ix CE ....................ix Reusing and Recycling ..............ix ® Bluetooth Safety and Compliance(9650) ........
Page 6
Zero Pressure ................. 19 Settings................... 20 Flow Setup ..................20 Workflows ..................21 Logging Profile ................21 Manage Data .................. 22 CHAPTER 4 SETTINGS ..............23 Display Order ................. 23 Actual/Standard ................24 Calibration ..................24 General Settings ................26 Unit Settings ...................
Safety This section provides instructions to ensure safe and proper operation ® of the VelociCalc Air Velocity Meter Series 9600. W A R N I N G S • The instrument must be used in the manner described in this manual.
C a u t i o n a n d W a r n i n g S y m b o l s The following symbols may accompany cautions and warnings to indicate the nature and consequences of hazards: Warns that uninsulated voltage within the instrument may have sufficient magnitude to cause electric shock.
7. Example 8. Example European symbol for non- disposable item. Item must be recycled. R o H S ® VelociCalc 9600 Series instruments are RoHS compliant. ® VelociCalc 9600 Series instruments are CE compliant. R e u s i n g a n d R e c y c l i n g Incorporated’s effort to have a minimal negative ®...
Page 10
NOTICE This device may not cause interference; this device must accept any interference, including interference that may cause undesired operation of the device. l’appareil ne doit pas produire de brouillage; l’appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est susceptible d’en compromettre le fonctionnement.
C H A P T E R 1 Unpacking and Parts Identification Carefully unpack the instrument and accessories from the shipping container. Check the individual parts against the list of components ® below. If anything is missing or damaged, notify TSI immediately.
O p t i o n a l P l u g I n P r o b e s Telescopic Thermoanemometer Probes Model No. Description Air Velocity and Temperature, Straight Probe Air Velocity and Temperature, Articulating Probe Air Velocity, Temperature, and Humidity, Straight Probe Air Velocity, Temperature, and Humidity, Articulating Probe Telescopic Thermohygrometer Probe Model No.
Optional Accessories and Replacement Parts Part No. Description Picture 800122 AC/DC Adapter/Power Supply 802241 USB Thermal Printer ® 801190 Bluetooth Printer (9650 only) 80211 Printer paper (5 rolls) 800681 IAQ probe stand 801748 Telescopic, articulated extension for 995 RV probe 372000000 8 ft.
Page 14
Part No. Description Picture 800130 Carrying case, small: Accommodates (1) meter and up to (2) probes (IAQ or thermoanemometer) ® ® VelociCalc /VeloCalc Pro Air Velocity Meter 9600 Series...
C a s e M a g n e t s , P r o b e H o l d e r a n d W r i s t S t r a p The back of the case features built-in magnets that allow for hands- free operation.
The telescoping probe contains the velocity, temperature, and humidity sensors. When using the probe, make sure the sensor window is fully exposed and the orientation dimple is facing upstream. N O T I C E For temperature and humidity measurements, make sure that at least 3 in.
Connecting the Static Pressure Probe The static pressure probe included with differential pressure capable meters is connected to the “+” port using the included tubing. The static pressure probe is used to measure the duct static pressure and features a magnet which holds the probe to the ductwork.
W A R N I N G DO NOT use the instrument or probes near hazardous voltage sources since serious injury could result. T h e r m o c o u p l e P o r t ® The 9600 Series VelociCalc Air Velocity Meter includes a thermocouple port at the base of the meter.
W A R N I N G S • Thermocouples from an alternate TSI ® supplier must have the metal sheath electrically isolated from the wires inside. Failure to meet this requirement may result in false readings, electrical shock, or fire hazard. •...
C o n n e c t i n g t o a C o m p u t e r ® Use the Computer Interface USB Cable provided with the VelociCalc meter to connect the instrument to a computer for downloading stored data.
Page 22
(This page intentionally left blank.) ® ® VelociCalc /VeloCalc Pro Air Velocity Meter 9600 Series...
C H A P T E R 3 Operational Overview K e y p a d B u t t o n N a m e s Button Names Button Image Programmable Soft Keys Main Menu Power Home (Dashboard) Navigation Keys Enter or OK I c o n s...
Temperature Relative humidity Dew point temperature Wet bulb temperature Air flow from thermoanemometer or rotating vane anemometer probe Differential pressure Velocity from pitot probe Airflow from pitot probe or K-factor Carbon dioxide Carbon monoxide Dashboard Icons Battery status AC power Manual Logging mode Continuous Logging mode Number of samples...
Soft Key Icons Save a sample Cancel sample measurement in process Closes a TestID and automatically increments to the next available TestID Display average measurements of current TestID Close display of average measurements of current TestID Progress indicator Go back to previous screen Advance to next screen Add a configuration Delete...
Zero differential pressure (9630, 9650, 9650-NB) Print Logging profile Flow setup Display order Duct traverse (9650, 9650-NB) D e f i n i t i o n s Out-of-Range An Out-of-Range error during the calibration of a sensor means the sensor’s offset or (Calibration) calibration slope adjustment has drifted outside ®...
L a n g u a g e S e l e c t i o n A language selection list is displayed the first time the meter is powered on. D a s h b o a r d Dashboard This is the main page for viewing live readings and logging data.
Air Density Setting Battery or AC Adapter Number of Samples Current Test ID Log Mode Measurements Display Average of current test ID Close Test ID NOTICE Measurements will only appear in the Dashboard page once they have been configured as visible in the Display Order page. A s s i g n P r o g r a m m a b l e S o f t K e y s The Dashboard offers a feature called Programmable Soft Keys.
M a i n M e n u Press the button to bring up the main menu. NOTICE The Main Menu items displayed are dependent on the meter model and connected probe. Zero Pressure will be displayed on Models 9630, 9650 and 9650-NB.
S e t t i n g s Select Settings from the Main Menu, to view the Settings page options. Refer to Chapter 4, Settings for detailed information about the device setting options. F l o w S e t u p Select Flow Setup to define duct sizes and choose a duct shape.
W o r k f l o w s Select Workflows from the Main Menu, to view the Workflow options. Refer to Chapter 7, Workflows for detailed information about the device setting options. Duct Traverse and Heat Flow are only available on the 9650 and 9650-NB.
M a n a g e D a t a Select Manage Data to display logged data stored in the device. Refer to Chapter 6, Manage Data for detailed information. ® ® VelociCalc /VeloCalc Pro Air Velocity Meter 9600 Series...
C H A P T E R 4 Settings Navigate to the Settings page by selecting Settings from the Main Menu. The Settings page options are: • Display Order • Actual/Standard • Calibration • General Settings • Unit Settings • Date and Time •...
NOTICES • The order of measurements on the Display Order page is the order shown on the Dashboard page. • Only those measurements configured as visible on the Display Order page are logged to a TESTID. A c t u a l / S t a n d a r d Select Actual/Standard to configure the settings used for velocity and flow measurements.
Page 35
Calibration Offset Adjustments Applicable Models Temperature Barometric Pressure Relative Humidity 960, 962, 9600, 9630, 9650, 964, 966, 980, 982, 995, 964, 966, 9650-NB 800220 980, 982, 800220 You can apply offset adjustments to temperature, relative humidity, and barometric pressure. The measurement displayed and logged will be the raw measurement added to the value...
A probe calibration collar (included with the 980 and 982 probes), zero calibration gas, span calibration gas, gas regulator and tubing are required to perform the calibration. The gas regulator used to control the flow should be capable of providing 0.3 L/min. Follow the on- screen instructions to complete the calibration, the procedure is the same for both carbon dioxide...
NOTICE Time Constant is the display averaging period. The display will update every second; however, the displayed reading will be the average over the time constant period. For example, if the time constant is 5 seconds, the display will update every second, but the displayed reading will be the average of the last 5 seconds.
D i s p l a y / P o w e r The Display/Power page is used to configure the following functions • Screen Brightness adjusts the brightness of the display. • Automatic Shutdown enables and disables automatic shutdown. If enabled, the meter will shut itself off after 20 minutes of inactivity.
C H A P T E R 5 Logging Profile and Custom TESTID Labels M a n u a l M o d e L o g g i n g To configure the meter to log when the Enter button is pressed, select Manual Mode.
C o n t i n u o u s M o d e L o g g i n g To configure the meter to log continuously, select Continuous Mode. The Time Constant (Sample Duration) setting determines how long all measurements are averaged for.
NO T IC E To scroll through the list of TESTIDs faster, press to page down and to page up through the list. C u s t o m i z e T E S T I D L a b e l s u s i n g T e s t I D . c s v ®...
Page 42
NOTICE TESTID labels are limited to 6 characters. Any additional characters will be discarded. ® ® VelociCalc /VeloCalc Pro Air Velocity Meter 9600 Series...
C H A P T E R 6 Manage Data From the Main Menu, select Manage Data. The Manage Data page contains all TESTIDs on the device. You can select a log file for viewing or deleting, as well as scroll through a list of log files.
V i e w i n g S a m p l e s To view Samples in a log file TESTID, navigate to the desired measurement in the Statistics page and press key. D e l e t e L o g F i l e s Select a log file then select the icon to delete it.
C H A P T E R 7 Workflows Workflows step you through the measurement process and assist in calculating and logging the measurement points. The following workflows are available: • Percent Outside Air (%OA) (9600, 9630, 9650, 9650-NB) • Heat Flow (9650, 9650-NB with 964 or 966 probe) •...
The Percent Outside Air feature offers the choice between using Temperature or CO for the Percent Outside Air study if a probe that supports either CO or Temperature is plugged in. Three measurements are necessary for Percent Outside Air calculations: Outside Air, Supply Air, and Return Air.
• Upstream measurements of Flow, Temperature, and Humidity are displayed. Press to capture these upstream measurements. o Press to accept the upstream readings o Press to retake upstream readings • After upstream measurement has been accepted, place probe downstream of the coil. o Press to capture these upstream measurements.
Page 48
After pressing to add a new duct size, you are brought to the Flow Setup page. To change the dimensions, use the arrow keys to navigate between Width and Height then press to begin editing the numbers by using the arrow keys. Press to save the new duct size.
Page 49
After selecting the traverse orientation (or after selecting a circular duct), the Insertion Depths screen will be shown. Note these insertion depths before proceeding. Press to proceed to the Duct Traverse working screen. Insert your probe into the first position indicated by the blue circle (traverse starts in the bottom left of the grid).
Page 50
(This page intentionally left blank.) ® ® VelociCalc /VeloCalc Pro Air Velocity Meter 9600 Series...
C H A P T E R 8 Maintenance C l e a n i n g / D i s i n f e c t i n g • ® Ensure the VelociCalc meter is turned off and not plugged into the AC/DC power supply.
B i - A n n u a l M a i n t e n a n c e C h e c k s • Make sure there are no cracks on the meter case. Cracks can create inconsistencies with how the electronics are supported inside the case, which can lead to damages.
C H A P T E R 9 Troubleshooting Table 1 lists the symptoms, possible causes, and recommended ® solutions for common problems encountered with the VelociCalc Velocity Meter. If your symptom is not listed, or if none of the ®...
Page 54
(This page intentionally left blank.) ® ® VelociCalc /VeloCalc Pro Air Velocity Meter 9600 Series...
A P P E N D I X A Specifications Specifications are subject to change without notice. 960 Thermoanemometer Straight Probe Velocity and Temperature Range: ..... 0 to 9999 ft/min (0 to 50 m/s), 0 to 200°F (-18 to 93°C) Accuracy: ....
Page 58
962 and 966 Thermoanemometer Probe Dimensions (962, 966) Probe Length: ....... 40 in. (101.6 cm) Probe Diameter of Tip: ....0.28 in. (7.0 mm) Probe Diameter of Base: ....0.51 in. (13.0 mm) Articulating Section Length: ..6.0 in. (15.24 cm) Diameter of Articulating Knuckle: .
Page 59
792 and 794 Thermocouple Probes Temperature Range: ..... -40 to 1200°F (-40 to 650°C) ±0.1% of reading +2°F Accuracy ....(±0.056% of reading +1.1°C) Resolution: ....0.1°F (0.1°C) Pitot Tubes (9630, 9650, 9650-NB) Range : ....250 to 15500 ft/min (1.27 to 78.7 m/s) Accuracy : ....
Page 60
Instrument Operating Conditions Altitude up to 4000 meters (only limited when plugged into the AC/DC adapter) Relative humidity up to 80% RH, non-condensing Pollution Degree II Overvoltage Category II Data Storage Capabilities Range: ......162,200 samples and 200 TESTIDs (one sample can contain multiple measurement types) Logging Interval Intervals: ......