Effects of Leakage Capacitance

Each measurement channel of the Analyzer is isolated from each other. However, these isolated
channels still have a leakage capacitance C
capacitance of the instrument itself, the isolation capacitance of the power adapter, and the anti-
interference capacitance inside the instrument. In general, the test object of the Analyzer has a higher
voltage relative to the earth, and we call this voltage common mode voltage V
voltage can generate a current i
certain impact on the measurement results of the Analyzer.
As shown in
Figure
A3, during voltage measurement, the current i
circuit through R
, R
s+
measurement and R
in
measurement voltage is:
the power supply under measurement R
terminal (black terminal) of the measurement circuit.
Figure A3. Effects of Leakage Current on Voltage Measurement
During current measurement, the current i
resistance of the test resistance and the common terminal of the measurement circuit R
voltage, that is coupled into the measurement circuit. The effect on the measurement
ΔI
[
(
=
i
R
current:
cs
Shunt
can cause measurement errors.
through the leakage capacitance C
cs
and R
. R
and R
s-
in
s+
is the test internal resistance of the power analyzer. The effect on the
ΔU
[
=
i
R
//(R
cs
S-
s-
Rs+
U
Rs-
Ics
cs
) R
]
+
R
, the voltage drop of i
S-
Shunt
relative to the earth, which consists of the distributed
s
are internal resistances of the power supply under
s-
)
]
+
R
. Because R
S+
in
has the main impact. R
Rin
2M Ohm
Cs
Vcom
∆U= Ics[Rs-//(Rs+ + Rin)]
flows through the test resistance R
cs
Power Analyzers
. The common mode
com
. This current i
s
cs
is coupled to the measurement
cs
is larger, internal resistance of
in
is connected to the common
s-
, and the internal
shunt
in R
and internal resistance R
Shunt
Appendix A
will have a
and forms a
s-
s -
81